Object: To assess the feasibility of measuring diffusion and perfusion fraction in vertebral bone marrow using the intravoxel incoherent motion (IVIM) approach and to compare two fitting methods, i.e., the non-negative least squares (NNLS) algorithm and the more commonly used Levenberg-Marquardt (LM) non-linear least squares algorithm, for the analysis of IVIM data.
Materials And Methods: MRI experiments were performed on fifteen healthy volunteers, with a diffusion-weighted echo-planar imaging (EPI) sequence at five different b-values (0, 50, 100, 200, 600 s/mm2), in combination with an STIR module to suppress the lipid signal. Diffusion signal decays in the first lumbar vertebra (L1) were fitted to a bi-exponential function using the LM algorithm and further analyzed with the NNLS algorithm to calculate the values of the apparent diffusion coefficient (ADC), pseudo-diffusion coefficient (D*) and perfusion fraction.
Results: The NNLS analysis revealed two diffusion components only in seven out of fifteen volunteers, with ADC=0.60±0.09 (10(-3) mm(2)/s), D*=28±9 (10(-3) mm2/s) and perfusion fraction=14%±6%. The values obtained by the LM bi-exponential fit were: ADC=0.45±0.27 (10(-3) mm2/s), D*=63±145 (10(-3) mm2/s) and perfusion fraction=27%±17%. Furthermore, the LM algorithm yielded values of perfusion fraction in cases where the decay was not bi-exponential, as assessed by NNLS analysis.
Conclusion: The IVIM approach allows for measuring diffusion and perfusion fraction in vertebral bone marrow; its reliability can be improved by using the NNLS, which identifies the diffusion decays that display a bi-exponential behavior.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.mri.2014.07.009 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!