Background: The hybrid tracer was introduced to complement intraoperative radiotracing towards the sentinel nodes (SNs) with fluorescence guidance.

Objective: Improve in vivo fluorescence-based SN identification for prostate cancer by optimising hybrid tracer preparation, injection technique, and fluorescence imaging hardware.

Design, Setting, And Participants: Forty patients with a Briganti nomogram-based risk >10% of lymph node (LN) metastases were included. After intraprostatic tracer injection, SN mapping was performed (lymphoscintigraphy and single-photon emission computed tomography with computed tomography (SPECT-CT)). In groups 1 and 2, SNs were pursued intraoperatively using a laparoscopic gamma probe followed by fluorescence imaging (FI). In group 3, SNs were initially located via FI. Compared with group 1, in groups 2 and 3, a new tracer formulation was introduced that had a reduced total injected volume (2.0 ml vs. 3.2 ml) but increased particle concentration. For groups 1 and 2, the Tricam SLII with D-Light C laparoscopic FI (LFI) system was used. In group 3, the LFI system was upgraded to an Image 1 HUB HD with D-Light P system.

Intervention: Hybrid tracer-based SN biopsy, extended pelvic lymph node dissection, and robot-assisted radical prostatectomy.

Outcome Measurements And Statistical Analysis: Number and location of the preoperatively identified SNs, in vivo fluorescence-based SN identification rate, tumour status of SNs and LNs, postoperative complications, and biochemical recurrence (BCR).

Results And Limitations: Mean fluorescence-based SN identification improved from 63.7% (group 1) to 85.2% and 93.5% for groups 2 and 3, respectively (p=0.012). No differences in postoperative complications were found. BCR occurred in three pN0 patients.

Conclusions: Stepwise optimisation of the hybrid tracer formulation and the LFI system led to a significant improvement in fluorescence-assisted SN identification. Preoperative SPECT-CT remained essential for guiding intraoperative SN localisation.

Patient Summary: Intraoperative fluorescence-based SN visualisation can be improved by enhancing the hybrid tracer formulation and laparoscopic fluorescence imaging system.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.eururo.2014.07.014DOI Listing

Publication Analysis

Top Keywords

hybrid tracer
16
fluorescence-based identification
12
fluorescence imaging
12
tracer formulation
12
lfi system
12
prostate cancer
8
vivo fluorescence-based
8
lymph node
8
computed tomography
8
postoperative complications
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!