Background: The aim of this study is to investigate the inter-scan reproducibility of kinetic parameters in atherosclerotic plaque using dynamic contrast-enhanced (DCE) cardiovascular magnetic resonance (CMR) in a multi-center setting at 3T.

Methods: Carotid arteries of 51 subjects from 15 sites were scanned twice within two weeks on 3T scanners using a previously described DCE-CMR protocol. Imaging data with protocol compliance and sufficient image quality were analyzed to generate kinetic parameters of vessel wall, expressed as transfer constant (K trans ) and plasma volume (v p ). The inter-scan reproducibility was evaluated using intra-class correlation coefficient (ICC) and coefficient of variation (CV). Power analysis was carried out to provide sample size estimations for future prospective study.

Results: Ten (19.6%) subjects were found to suffer from protocol violation, and another 6 (11.8%) had poor image quality (n=6) in at least one scan. In the 35 (68.6%) subjects with complete data, the ICCs of K trans and v p were 0.65 and 0.28, respectively. The CVs were 25% and 62%, respectively. The ICC and CV for v p improved to 0.73 and 28% in larger lesions with analyzed area larger than 25 mm2. Power analysis based on the measured CV showed that 50 subjects per arm are sufficient to detect a 20% difference in change of K trans over time between treatment arms with 80% power without consideration of the dropout rate.

Conclusion: The result of this study indicates that quantitative measurement from DCE-CMR is feasible to detect changes with a relatively modest sample size in a prospective multi-center study despite the limitations. The relative high dropout rate suggested the critical needs for intensive operator training, optimized imaging protocol, and strict quality control in future studies.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4237824PMC
http://dx.doi.org/10.1186/s12968-014-0051-7DOI Listing

Publication Analysis

Top Keywords

atherosclerotic plaque
8
plaque dynamic
8
dynamic contrast-enhanced
8
cmr multi-center
8
multi-center study
8
inter-scan reproducibility
8
kinetic parameters
8
image quality
8
power analysis
8
sample size
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!