Forces in spinal cannulation and breaches ex vivo.

Proc Inst Mech Eng H

Biomedical Engineering, University of Calgary, Calgary, AB, Canada McCaig Institute for Bone and Joint Health, University of Calgary, Calgary, AB, Canada Department of Civil Engineering, University of Calgary, Calgary, AB, Canada

Published: July 2014

Background: Pedicle screw insertion, to stabilize or correct the spine, relies on creating a probe path with the correct trajectory to prevent unsafe breaching of the cortical wall. Safe pedicle cannulation is aided when the surgeon can feel the difference between a safe and unsafe path. Pedicle probe forces and torques are currently unknown. The purpose of this study was to investigate the forces and torques encountered while cannulating the pedicle tract in both correct and incorrect cannulations.

Methods: Two experienced surgeons used a standard lumbar probe modified to incorporate a 6 degree-of-freedom load cell to cannulate and breach the T12 to S1 vertebrae of six fresh frozen cadavers (3 males, 3 females, ages 65 to 92). A total of 76 pedicles were tested.

Findings: Cannulation axial forces averaged 48 N (standard deviation = 13 N), medial breach 129 N (standard deviation = 25 N), and lateral breach 86 N (standard deviation = 27 N). Cannulation values were significantly lower than the breach values in all 6 degrees of freedom (p < 0.001). There were significant differences between specimens, including males and females, and between degrees of freedom, but no significant right and left differences or by vertebral level.

Conclusion: A large range of cannulation and breach forces and torques were measured due to variations in bone quality and geometry, as experienced clinically. This is the first time that the absolute and relative force and torque levels have been reported, to our knowledge.

Download full-text PDF

Source
http://dx.doi.org/10.1177/0954411914540632DOI Listing

Publication Analysis

Top Keywords

forces torques
12
standard deviation
12
males females
8
degrees freedom
8
forces
5
cannulation
5
breach
5
forces spinal
4
spinal cannulation
4
cannulation breaches
4

Similar Publications

This study aims to compare the sinking and shifting of an enhanced escape-proof intervertebral fusion device with a traditional TLIF intervertebral fusion device.Five specimens each of the improved escape-resistant intervertebral cage and the traditional TLIF cage were selected. Four types of mechanical tests were conducted on each cage, Furthermore, a blade-cutting torque test was performed on the escape-resistant cage, with the recording of load-displacement curves and mechanical values.

View Article and Find Full Text PDF

Background: Older adults with cancer are vulnerable to declines in muscle performance (e.g., strength, speed, duration of muscular contraction), which are associated with worse cancer-related outcomes.

View Article and Find Full Text PDF

Background And Purpose: Muscle strength, power, and mass decline with aging, leading to functional loss highly correlated with balance and falls in older adults. Lower limb muscle function is critical for fall prevention in older adults, and hip abductor force and rapid force development have been shown to be important during stepping tasks. However, it remains unclear whether hip abductor muscle function changes with aging.

View Article and Find Full Text PDF

Nanostructured dielectric metasurfaces offer unprecedented opportunities to control light-matter momentum exchange, and thereby the forces and torques that light can exert on matter. Here we introduce optical metasurfaces as components of ultracompact untethered microscopic metaspinners capable of efficient light-induced rotation in a liquid environment. Illuminated by weakly focused light, a metaspinner generates torque via photon recoil through the metasurfaces' ability to bend light towards high angles despite their sub-wavelength thickness, thereby creating orbital angular momentum.

View Article and Find Full Text PDF

This article presents a laboratory device by which the course of two signals can be detected using two types of sensors-strain gauges and the DEWESoft DS-NET measuring apparatus. The values of the coefficient of friction of the brake lining when moving against the rotating shell of the brake drum were determined from the physical quantities sensed by tensometric sensors and transformed into electrical quantities. The friction coefficient of the brake lining on the circumference of the rotating brake disc shell can be calculated from the known values measured by the sensors, the design dimensions of the brake, and the revolutions of the rotating parts system.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!