Background: Simultaneous multipolar ablation catheters have been proposed to simplify pulmonary vein isolation (PVI) in paroxysmal atrial fibrillation (AF). Recently, a new multipolar irrigated radiofrequency (RF) ablation catheter (nMARQ™, Biosense Webster Inc., Diamond Bar, CA, USA) combining both 3-dimensional electroanatomic mapping and multipolar open-irrigated ablation capability has been developed. Aim of our study was to assess feasibility, acute and short-term success and safety of PVI by the use of this new technology with particular regard to the incidence of postablation silent cerebral ischemia (SCI).
Methods And Results: Twenty-five patients (76% males; age 57 ± 13 years) with paroxysmal AF underwent PVI using the nMARQ™ catheter. PVI, confirmed by Lasso catheter mapping, was achieved in 100 out of 102 pulmonary veins (98%) identified, and final PVI was obtained in 24 out of 25 (96%) patients. The overall concordance between Lasso and nMARQ™ signals in demonstrating PVI was 78%. No major procedural complications occurred and no patient suffered SCI, on the basis of cerebral magnetic resonance imaging performed before and after the procedure. Following a 6-month follow-up, 17/25 (68%) patients remained free from AF without antiarrhythmic drugs.
Conclusions: In our preliminary experience, PVI with nMARQ™ catheter appears to be feasible and safe, without incidence of SCI. Long-term clinical efficacy has to be evaluated in further studies.