Objective: To evaluate the fresh gas flow (FGF) rate requirements for the Humphrey ADE semi-closed breathing system in the Mapleson A mode; to determine the FGF at which rebreathing occurs, and compare the efficiency of this system to the Bain (Mapleson D) system in spontaneously breathing cats and small dogs.
Study Design: Prospective clinical study.
Animals: Twenty-five healthy (ASA score I or II) client-owned cats and dogs (mean ± SD age 4.7 ± 5.0 years, and body weight 5.64 ± 3.26 kg) undergoing elective surgery or minor procedures.
Methods: Anaesthesia was maintained with isoflurane delivered via the Humphrey ADE system in the A mode using an oxygen FGF of 100 mL kg(-1) minute(-1). The FGF was then reduced incrementally by 5-10 mL kg(-1) minute(-1) at approximately five-minute intervals, until rebreathing (inspired CO(2) >5 mmHg (0.7 kPa)) was observed, after which flow rates were increased. In six animals, once the minimum FGF at which rebreathing occurred was found, the breathing system was changed to the Bain, and the effects of this FGF delivery examined, before FGF was increased.
Results: Rebreathing did not occur at the FGF recommended by the manufacturer for the ADE. The mean ± SD FGF that resulted in rebreathing was 60 ± 20 mL kg(-1) minute(-1). The mean minimum FGF at which rebreathing did not occur with the ADE was 87 ± 39 mL kg(-1) minute(-1). This FGF resulted in significant rebreathing (inspired CO(2) 8.8 ± 2.6 mmHg (1.2 ± 0.3 kPa)) on the Bain system.
Conclusions: The FGF rates recommended for the Humphrey ADE are adequate to prevent rebreathing in spontaneously breathing cats and dogs <15 kg.
Clinical Relevance: The Humphrey ADE system used in the A mode is a more efficient alternative to the Bain system, for maintenance of gaseous anaesthesia in spontaneously breathing cats and small dogs.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/vaa.12209 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!