Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Rainbow Pulse CO-Oximetry technology (Masimo Corporation, Irvine, CA) provides continuous and noninvasive measurement of arterial hemoglobin concentration (SpHb). We assessed the trending and accuracy of SpHb by this innovative monitoring compared with Hb concentration obtained with conventional laboratory techniques (Hb) in children undergoing surgical procedures with potential for substantial blood loss.
Methods: Hb concentrations were recorded from Pulse CO-Oximetry and a conventional hematology analyzer. Regression analysis and 4-quadrant plot were used to evaluate the trending for changes in SpHb and Hb measurements (ΔSpHb and ΔHb). Bias, precision, and limits of agreement of SpHb and of in vivo adjusted SpHb (SpHb - first bias to HB) compared with Hb were calculated.
Results: One hundred fifty-eight SpHb-Hb data pairs and 105 delta pairs (ΔSpHb and ΔHb) from 46 patients aged 2 months to 17 years with Hb ranging from 16.7 to 7.9 g/dL were collected. To evaluate trending, the delta pairs (ΔSpHb and ΔHb) were plotted, which revealed a positive correlation (ΔSpHb = 0.022 + 0.76ΔHb) with correlation coefficient r = 0.76, 95% CI [confidence interval] = 0.57-0.86. The bias and precision of SpHb to Hb and in vivo adjusted SpHb were 0.4 ± 1.3 g/dL and 0.1 ± 1.2 g/dL, respectively; the limits of agreement were -2.0 to 3.2 g/dL before in vivo adjustment and -2.4 to 2.2 g/dL after in vivo adjustment (P value = 0.04). The mean percent bias (from the reference Hb concentration) decreased from 4.1% ± 11.9% to 0.7% ± 11.3% (P value = 0.01). No drift in bias over time was observed during the study procedure. Of patient demographic and physiological factors tested for correlation with the SpHb, only perfusion index at sensor site showed a weak correlation.
Conclusions: The accuracy of SpHb in children with normal Hb and mild anemia is similar to that previously reported in adults and is independent of patient demographic and physiological states except for a weak correlation with perfusion index. The trending of SpHb and Hb in children with normal Hb and mild anemia showed a positive correlation. Further studies are necessary in children with moderate and severe anemia.