Background: Methods to improve gap balancing in total knee arthroplasty (TKA) include the development of calibrated distractors and various devices to determine the distances of the gaps. However, few studies have validated the accuracy or precision of computer navigation to determine these measurements, especially gaps created after bone cuts have been made; doing so would be important, because optimal surgical technique relies on appropriate gap spacing.

Questions/purposes: We investigated the ability of a new image-free computer navigation surface registration protocol to measure gap distances in TKA.

Methods: Eight embalmed cadaveric specimens of the lower extremity were used. A surface registration software protocol defined the most distal and posterior surface points of the femoral condyles and the navigation system measured the distance of the most distal femoral condyle point to the surface of the tibia after tibial resection. The tibial resection was perpendicular to the mechanical axis and was cut with a 7° posterior slope. The navigation system measured gaps spaced by modular spacing blocks at 5° intervals from full extension to 120° of flexion. Repeatability assessed repeated measures by one surgeon. Reproducibility was assessed by performing the same measurements after complete reregistration of the computer protocol to the cadaver bones.

Results: The gaps measured by the computer were statistically the same as those assessed with the use of blocks with a maximum measurement error of 1 mm. Reregistration did introduce error into the measurement. The gaps changed with position of knee flexion, and there was gradual and significant stretching of the gaps with repeated measurements.

Conclusions: Preliminary testing shows that computer navigation can reproduce static measurements reliably and with equal accuracy as spacer blocks. We have not demonstrated that this could be applied in a dynamic setting.

Clinical Relevance: This computer navigation system has sufficient precision to warrant investigation in the clinical setting for measuring gaps created during the surgical procedure.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4390933PMC
http://dx.doi.org/10.1007/s11999-014-3785-5DOI Listing

Publication Analysis

Top Keywords

computer navigation
16
navigation system
12
gaps created
8
surface registration
8
system measured
8
tibial resection
8
gaps
7
computer
6
navigation
6
precise computer-navigated
4

Similar Publications

The state-of-the-art approach to open reduction and fixation (ORIF) of zygoma fracture fragments is based on manual skills. Achieving high accuracy can be challenging. Our feasibility study on deceased body donors with artificial zygomatic fractures investigated whether virtual repositioning of the fractures and the use of customised 3D-printed titanium osteosynthesis plates was similar in accuracy to the conventional manual procedure, and whether the method was applicable in a clinical setting.

View Article and Find Full Text PDF

Finding of situs inversus in endoscopic sleeve gastroplasty: Case report.

Int J Surg Case Rep

December 2024

IFSO, USA; Fundación Santa Fe de Bogotá, Colombia.

Introduction And Importance: Situs inversus is an anatomical rare condition in which visceral organs are not located in its normal position, with a reversal anatomical orientation.

Case Presentation: We present a case of an 27-year-old male with a Body Mass Index (BMI) of 36.02 Kg/m2, who was programed for a Endoscopic Sleeve Gastroplasty (ESG), in which Situs inversus was documented.

View Article and Find Full Text PDF

Background: Complete mesocolon excision (CME) and central vascular ligation for right colonic cancers have been developed to improve oncological outcomes. However, it has been linked with a higher risk of morbidity and technical difficulties in operating near major vessels. This study investigated the impact of preoperative surgical planning utilizing CT reconstruction on surgical outcomes in right colectomy with CME.

View Article and Find Full Text PDF

Small-scale continuum robots hold promise for interventional diagnosis and treatment, yet existing models struggle to achieve small size, precise steering, and visualized functional treatment simultaneously, termed an "impossible trinity". This study introduces an optical fiber-based continuum robot integrated imaging, high-precision motion, and multifunctional operation abilities at submillimeter-scale. With a slim profile of 0.

View Article and Find Full Text PDF

Real-Time Tractography-Assisted Neuronavigation for Transcranial Magnetic Stimulation.

Hum Brain Mapp

January 2025

Department of Neuroscience and Biomedical Engineering, School of Science, Aalto University, Espoo, Finland.

State-of-the-art navigated transcranial magnetic stimulation (nTMS) systems can display the TMS coil position relative to the structural magnetic resonance image (MRI) of the subject's brain and calculate the induced electric field. However, the local effect of TMS propagates via the white-matter network to different areas of the brain, and currently there is no commercial or research neuronavigation system that can highlight in real time the brain's structural connections during TMS. This lack of real-time visualization may overlook critical inter-individual differences in brain connectivity and does not provide the opportunity to target brain networks.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!