Purpose: This study was conducted to directly compare the high-resolution research tomograph (HRRT) (high-resolution brain) and HR+ (standard whole-body) positron emission tomography (PET) only scanners for quantitative brain studies using three tracers with vastly different tracer distributions.
Procedures: Healthy volunteers underwent successive scans on HR+ and HRRT scanners (in random order) using either (R)-[(11)C]verapamil (n = 6), [(11)C]raclopride (n = 7) or [(11)C]flumazenil (n = 7). For all tracers, metabolite-corrected plasma-input functions were generated.
Results: After resolution matching, HRRT-derived kinetic parameter values correlated well with those of HR+ for all tracers (intraclass correlation coefficients ≥0.78), having a good absolute interscanner test-retest variability (≤15 %). However, systematic differences can be seen for HRRT-derived kinetic parameter values (range -13 to +15 %).
Conclusion: Quantification of kinetic parameters based on plasma-input models leads to comparable results when spatial resolution between HRRT and HR+ data is matched. When using reference-tissue models, differences remain that are likely caused by differences in attenuation and scatter corrections and/or image reconstruction.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s11307-014-0766-8 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!