Background: In this study, we assessed the geometric accuracy of an automated positioning system in Gamma Knife (GK) surgery. Specifically, we looked at the total spatial uncertainty over the entire treatment range of GK stereotactic radiosurgery (SRS) procedures in both the GK model C and the Perfexion (PFX).

Methods: An originally-developed phantom and a radiochromic film were used for obtaining actual dose distributions. The phantom, with inserted films on different axial planes (z = 60, 75, 100, 125, 140 mm), sagittal planes (x = 60, 75, 100, 125, 140 mm), and coronal planes (y = 60, 75, 100, 125, 140 mm), was placed on a Leksell skull frame. Computed tomography (CT) was then performed with a stereotactic localizer box attached to the frame, and dose planning was made using the Leksell GammaPlan treatment planning system. The phantom finally received beam delivery using a single shot of a 4-mm collimator helmet. The discrepancy between the planned shot position and the irradiated center position was evaluated by a dedicated film analysis software.

Results: The total uncertainty of CT-based GK SRS was less than 1 mm for almost all measured points over the stereotactic space in both the model C and the PFX. In addition, the geometric accuracy of the automated positioning system was estimated to be less than 0.1 mm and equal to 0.5 mm in the central and peripheral areas, respectively.

Conclusions: We confirmed that the total spatial uncertainties of both the GK model C and the PFX are acceptable for clinical use.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00701-014-2178-6DOI Listing

Publication Analysis

Top Keywords

automated positioning
12
positioning system
12
100 125
12
125 140
12
spatial uncertainty
8
gamma knife
8
stereotactic radiosurgery
8
geometric accuracy
8
accuracy automated
8
total spatial
8

Similar Publications

Purpose: During endovascular revascularization interventions for peripheral arterial disease, the standard modality of X-ray fluoroscopy (XRF) used for image guidance is limited in visualizing distal segments of infrapopliteal vessels. To enhance visualization of arteries, an image registration technique was developed to align pre-acquired computed tomography (CT) angiography images and to create fusion images highlighting arteries of interest.

Methods: X-ray image metadata capturing the position of the X-ray gantry initializes a multiscale iterative optimization process, which uses a local-variance masked normalized cross-correlation loss to rigidly align a digitally reconstructed radiograph (DRR) of the CT dataset with the target X-ray, using the edges of the fibula and tibia as the basis for alignment.

View Article and Find Full Text PDF

Objective: This retrospective study aimed to evaluate morphometric changes in mandibular condyles of patients with skeletal Class III malocclusion following two-jaw orthognathic surgery planned using virtual surgical planning (VSP) and analysed with automated three-dimensional (3D) image analysis based on deep-learning techniques.

Materials And Methods: Pre-operative (T1) and 12-18 months post-operative (T2) Cone-Beam Computed Tomography (CBCT) scans of 17 patients (mean age: 24.8 ± 3.

View Article and Find Full Text PDF

Being part of a social structure offers chances for social learning vital for survival and reproduction. Nevertheless, studying the neural mechanisms of social learning under laboratory conditions remains challenging. To investigate the impact of socially transmitted information about rewards on individual behavior, we used Eco-HAB, an automated system monitoring the voluntary behavior of group-housed mice under seminaturalistic conditions.

View Article and Find Full Text PDF

Robotic sugarbaker parastomal hernia repair: updated series and outcomes.

Hernia

January 2025

Division of Gastrointestinal and Minimally Invasive Surgery, Department of Surgery, Carolinas Medical Center, 1025 Morehead Medical Drive Suite 300, Charlotte, NC, 28204, USA.

Purpose: To present updated outcomes after previously describing a novel technique for the robotic repair of parastomal hernias.

Methods: Patients who underwent parastomal hernia repair with a robotic Sugarbaker technique at a tertiary hernia center were identified from an institutional database. The approach involves mesh placement in the intraperitoneal or preperitoneal position after closure of the fascial defect.

View Article and Find Full Text PDF

Accuracy of robot-assisted implant surgery versus freehand placement: a retrospective clinical study.

Int J Implant Dent

January 2025

Key Laboratory of Shaanxi Province for Craniofacial Precision Medicine Research, College of Stomatology, Xi'an Jiaotong University, Xi'an, China.

Purpose: This study evaluated the accuracy of implant placement using a robotic system (Remebot) compared to freehand surgery and explored factors influencing accuracy.

Methods: This retrospective study included 95 implants placed in 65 patients, divided into robot-assisted (50 implants) and freehand (45 implants) groups. Platform, apical, and angular deviations were measured by superimposing preoperative plans and the postoperative CBCT images.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!