Background: Benzodiazepines are widely prescribed, and compliance of patients with benzodiazepine therapy is often monitored using urine specimens. Although various commercially available benzodiazepines immunoassays are widely used for compliance monitoring, such immunoassays usually have low cross-reactivity with glucuronide metabolites. We studied the effect of hydrolyzing such glucuronide before analysis to reevaluate suitability of Enzyme multiplied immunoassay technique benzodiazepine immunoassay for monitoring compliance with benzodiazepine therapy.

Methods: In 31 urine specimens collected from patients taking benzodiazepines, the true analyte concentrations were determined (after hydrolyzing glucuronide metabolites using beta-glucuronidase) using liquid chromatography-tandem mass spectrometry. These urine specimens were reanalyzed using EMIT benzodiazepine assay (Flex Reagent Cartridge; Siemens Diagnostics) and Vista analyzer.

Results: We observed false negative test results with EMIT in 11 of 31 specimens analyzed where liquid chromatography-tandem mass spectrometry values were above the 200 ng/mL cutoff concentration, but EMIT benzodiazepine assay showed a negative result, indicating that despite hydrolysis of the specimen to liberate parent drug (glucuronide metabolite often has poor cross-reactivity), the false negative rate using EMIT assay was 35.5%.

Conclusions: Patient compliance with benzodiazepine therapy must be monitored using a chromatographic method.

Download full-text PDF

Source
http://dx.doi.org/10.1097/FTD.0000000000000117DOI Listing

Publication Analysis

Top Keywords

emit benzodiazepine
12
benzodiazepine therapy
12
hydrolyzing glucuronide
12
glucuronide metabolites
12
urine specimens
12
benzodiazepine
8
benzodiazepine immunoassay
8
immunoassay monitoring
8
monitoring compliance
8
compliance patients
8

Similar Publications

Trigeminal neuralgia (TN) is a severe form of neuropathic pain frequently associated with anxiety. The chronic constriction injury of the infraorbital nerve (CCI-ION) of rodents is a well-established model to study sensory alterations related to TN. However, few studies have addressed the emotional component of pain, which is fundamental to increase its translational capability.

View Article and Find Full Text PDF

Ultrasonic vocalization (USVs) is a promising tool to measure behavioral anxiety in rodents as USV recording is noninvasive, behaviorally relevant, ethological, and reproducible. Studies reporting the effects of stress-induced USVs in adult mice remain limited and debated. We investigated the conditions under which mice emit aversive USVs and evaluated the effects of psychiatric drugs on stress-induced USVs.

View Article and Find Full Text PDF

Objectives: The study aimed to develop a validated LC-MS / MS method for the measurement of carbamazepine and carbamazepine-10,11-epoxide (CBZE) levels, to compare the carbamazepine concentrations measured by AbSciex API 3200 LC-MS/MS and Beckman Coulter Emit® 2000 immunoassay and to investigate the effect of carbamazepine concentrations on various hematological and biochemical parameters.

Methods: For the measurement of carbamazepine and CBZE levels, a validated LC-MS / MS method was developed. Serum carbamazepine levels of patients were measured by AbSciex API 3200 LC-MS / MS and Beckman Coulter Emit® 2000 immunoassay.

View Article and Find Full Text PDF

Comparison of several immunoassays used in drugs of abuse screening: Assessment against gold standard methods and calculation of measurement uncertainty.

J Pharmacol Toxicol Methods

November 2020

NSW Health Pathology, Forensic & Analytical Science Service (FASS), Sydney, Australia; Faculty of Medicine and Health, Sydney University, Australia. Electronic address:

Background: Immunoassays provide simple, powerful and inexpensive screening methods for urine drug screening. Other substances and/or factors may interfere with the test and cause false or positive results. It is essential to understand the differences between methods to be able to evaluate their impact on the results.

View Article and Find Full Text PDF

Infection of toxoplasma gondii (TG), an intracellular neurotropic parasitic protozoon, has been associated with various neuropsychiatric disorders. TG is usually diagnosed from serological sample in which a positive test for Anti-TG immunoglobulin G (IgG) indicates TG infection (toxoplasmosis). The research was conducted to test the hypothesis that TG infection may be associated with substance abuse.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!