Background: Hemolysis, icterus, and lipemia (HIL) cause preanalytical interference and vary unpredictably with different analytical equipments and measurement methods. We developed an integrated reporting system for verifying HIL status in order to identify the extent of interference by HIL on clinical chemistry results.
Methods: HIL interference data from 30 chemical analytes were provided by the manufacturers and were used to generate a table of clinically relevant interference values that indicated the extent of bias at specific index values (alert index values). The HIL results generated by the Vista 1500 system (Siemens Healthcare Diagnostics, USA), Advia 2400 system (Siemens Healthcare Diagnostics), and Modular DPE system (Roche Diagnostics, Switzerland) were analyzed and displayed on physicians' personal computers.
Results: Analytes 11 and 29 among the 30 chemical analytes were affected by interference due to hemolysis, when measured using the Vista and Modular systems, respectively. The hemolysis alert indices for the Vista and Modular systems were 0.1-25.8% and 0.1-64.7%, respectively. The alert indices for icterus and lipemia were <1.4% and 0.7% in the Vista system and 0.7% and 1.0% in the Modular system, respectively.
Conclusions: The HIL alert index values for chemical analytes varied depending on the chemistry analyzer. This integrated HIL reporting system provides an effective screening tool for verifying specimen quality with regard to HIL and simplifies the laboratory workflow.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4071188 | PMC |
http://dx.doi.org/10.3343/alm.2014.34.4.307 | DOI Listing |
Clin Chem Lab Med
November 2024
Utrecht University, Utrecht, The Netherlands.
Objectives: This study examined the comparability of venous and capillary blood samples with regard to routine chemistry analytes.
Methods: Venous and capillary blood samples were collected from adult patients to assess comparability of alanine transaminase, albumin, alkaline phosphatase, apolipoprotein B, aspartate aminotransferase, total bilirubin, calcium, chloride, creatin kinase, creatinine, C-reactive protein, ferritin, folic acid, free T4, gamma glutamyltransferase, glucose, high density lipoprotein cholesterol, iron, lipase, lipoprotein a, magnesium, phosphate, postassium, prostate specific antigen, sodium, total cholesterol, total protein, transferrin, triglycerides, thyroid stimulating hormone, urate, urea, vitamin B12 and 25-hydroxyvitamin-D3. Furthermore, hemolysis-icterus-lipemia Index (HIL-Index) was measured for all samples.
Scand J Clin Lab Invest
November 2024
Department of Experimental Medicine, University of Rome Tor Vergata, Rome, Italy.
The erythrocyte sedimentation rate (ESR) is a widely used diagnostic test, influenced by all physiological and pathological conditions that can bias blood rheology by interfering factors. This study aimed to evaluate the performance of the CUBE 30 touch ESR analyzer in samples with preanalytical variables, as lipemia, hemolysis, and icterus or in presence of fibrinogen., Moreover we focused to define the maximum time limits to ensure a reliable ESR measure.
View Article and Find Full Text PDFCureus
July 2024
Pediatrics and Child Health, Jawaharlal Nehru Medical College, Datta Meghe Institute of Higher Education and Research, Wardha, IND.
Red cell pyruvate kinase (PK) deficiency is a genetic disorder affecting the enzyme PK in red blood cells. A deficiency in PK leads to hemolytic anemia. Hypertriglyceridemia means elevated levels of triglycerides in the blood.
View Article and Find Full Text PDFClin Chim Acta
January 2025
Department of Laboratory Medicine, Wonju Severance Christian Hospital, Yonsei University Wonju College of Medicine, Wonju, Korea. Electronic address:
Objective: Delta bilirubin (albumin-covalently bound bilirubin) may provide important clinical utility in identifying impaired hepatic excretion of conjugated bilirubin, but it cannot be measured in real-time for diagnostic purposes in clinical laboratories.
Methods: A total of 210 samples were collected, and their delta bilirubin levels were measured four times using high-performance liquid chromatography. Data collected included age, sex, diagnosis code, delta bilirubin, total bilirubin, direct bilirubin, total protein, albumin, globulin, aspartate aminotransferase, alanine transaminase, alkaline phosphatase, gamma-glutamyl transferase, lactate dehydrogenase, hemoglobin, serum hemolysis value, hemolysis index, icterus value (Iv), icterus index (Ii), lipemia value (Lv), and lipemia index.
Ann Clin Lab Sci
May 2024
Department of Pathology and Laboratory Medicine, Pennsylvania State University College of Medicine, Hershey, PA, USA
Objective: We validated an automated microfluidic interleukin-6 (IL-6) immunoassay on the Ella platform for clinical use.
Methods: The assay was validated for precision, lower limit of quantification, analytical measurement range, accuracy, specificity, interference of biotin, tocilizumab, GP130, IL-6R, hemolysis, icterus, and lipemia, and establishing the reference value. The clinical performance was evaluated in 96 COVID-19 patients.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!