Background: The purpose of this study is to design and evaluate a new reduced scan time three-dimensional (3D) FLuid Attenuated Inversion Recovery (FLAIR) sequence.
Methods: The 3D FLAIR sequence was modified so that the repetition time was modulated in a predetermined smooth manner (3D mFLAIR). Inversion times were adjusted accordingly to maintain cerebrospinal fluid (CSF) suppression. Simulations were performed to determine SNR for gray matter (GM), white matter (WM), and CSF. Fourteen volunteers were imaged using the modified and product sequence. SNR measurements were performed in GM, WM, and CSF. Mean value and the 95% confidence interval ([CI]) were assessed. Scan time for the 3D FLAIR and 3D mFLAIR sequences was measured.
Results: There was no statistically significant difference in the SNR measured in GM (P value = 0.5; mean SNR = 42.8 [CI]: 38.2-45.5 versus 42.2 [CI]: 38.3-46.1 for 3D FLAIR and 3D mFLAIR, respectively) and WM (P value = 0.25; mean SNR = 32.1 [CI]: 30.3-33.8 versus 32.9 [CI]: 31.1-34.7). Scan time reduction greater than 30% was achieved for the given parameter set with the 3D mFLAIR sequence.
Conclusion: Scan time for 3D FLAIR can be effectively reduced by modulating repetition and inversion time in a predetermined manner while maintaining the SNR and CNR of a constant TR sequence.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4587400 | PMC |
http://dx.doi.org/10.1002/jmri.24679 | DOI Listing |
Sensors (Basel)
December 2024
State Key Laboratory of Robotics and System, Harbin Institute of Technology (HIT), Harbin 150001, China.
Some large social environments are expected to use Covered Path Planning (CPP) methods to handle daily tasks such as cleaning and disinfection. These environments are usually large in scale, chaotic in structure, and contain many obstacles. The proposed method is based on the improved SCAN-STC (Spanning Tree Coverage) method and significantly reduces the solution time by optimizing the backtracking module of the algorithm.
View Article and Find Full Text PDFPharmaceuticals (Basel)
December 2024
Department of Nuclear Medicine, First Medical Center, Chinese PLA General Hospital, Fuxing Road 28, Beijing 100853, China.
Objective: Currently, Ga-labeled somatostatin analogs (SSAs) are the most commonly used imaging agents for patients with neuroendocrine tumors (NETs) in clinical practice, demonstrating good results in tumor diagnosis. For applications in peptide receptor radionuclide therapy (PRRT), targeted drugs should have high tumor uptake and prolonged tumor retention time. To enhance the uptake and retention of tracers in NETs, our goal is to design a Ga-labeled heterodimer for optimizing pharmacokinetics and assess whether this form is more efficacious than its monomeric equivalents.
View Article and Find Full Text PDFMedicina (Kaunas)
November 2024
Department of Clinical and Experimental Medicine, University of Foggia, 71122 Foggia, Italy.
The purpose of this case report is to examine the management of vestibular bone fenestration during alveolar socket preservation using the Periosteal Inhibition (PI) approach. Here, for the first time, the PI technique, which has been shown to be successful in maintaining intact cortical bone, is examined in the context of a bone defect. : After an atraumatic extraction of a damaged tooth, a vestibular bone fenestration was discovered in the 62-year-old male patient.
View Article and Find Full Text PDFJ Clin Med
December 2024
Department of Anaesthesiology & Intensive Care Medicine, Klinikum Rechts der Isar, School of Medicine and Health, Technical University Munich, 81675 Munich, Germany.
: Skeletal muscle mass depletion adversely affects critically ill patient outcomes. Standardized methods for assessing muscle mass in this population are limited, particularly regarding changes during ICU stays and their implications for risk stratification. : In this secondary analysis of our prospective data registry of surgical ICU patients, we used a single slice extracted from a computed tomography scan to determine the patient's direction of absolute change in skeletal muscle mass between two different time points (-14 d to +0 d and +5 d to +21 d) during his or her critical illness.
View Article and Find Full Text PDFJ Clin Med
December 2024
University Hospital of Grenoble Alpes, INSERM U1300, 38043 Grenoble, France.
Cardiac implantable electronic devices and their integrated thoracic impedance sensors have been used to detect sleep apnea for over a decade now. Despite their usage in daily clinical practice, there are only limited data on their diagnostic accuracy. AIRLESS and UPGRADE were prospective investigator-driven trials meant to validate the AP scan (Boston Scientific, Marlborough, MA, USA) in heart failure cohorts.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!