Purpose: Magnetization transfer (MT) MRI can be effective for the diagnosis of a broad range of fibrotic diseases, including liver fibrosis. However, respiratory motion, a major source of artifacts in thoracic and abdominal MR imaging, can obscure important anatomic structures, making diagnosis difficult. In this study, we explored the potential to combine free-breathing (FB) respiratory self-gating (RSG) methods with MT saturation for FB MT ratio (MTR) measurements of abdominal organs.

Methods: A respiratory self-gated multiple-gradient recalled echo sequence with MT presaturation (RSG-MT GRE) was developed and applied in a series of seven normal volunteers. We compared the MTR values of liver, pancreas, kidney, spleen, and posterior paraspinal muscle measured using our RSG-MT GRE sequence and a conventional MT GRE sequence.

Results: RSG consistently reduced motion artifacts within MT-weighted images acquired during FB, improved the accuracy of FB MTR measurements, and produced comparable MTRs to breath-holding MTR measurements.

Conclusion: RSG approaches may offer to improve the utility of MT-weighted imaging methods for the assessment of fibrotic diseases and tumor desmoplasia in abdominal organs.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4276560PMC
http://dx.doi.org/10.1002/mrm.25341DOI Listing

Publication Analysis

Top Keywords

respiratory self-gating
8
magnetization transfer
8
transfer mri
8
fibrotic diseases
8
mtr measurements
8
rsg-mt gre
8
respiratory
4
self-gating free-breathing
4
free-breathing magnetization
4
mri abdomen
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!