Purpose: To investigate visual outcomes of Visian implantable Collamer lens (ICL) implantation in eyes with a shallow anterior chamber depth (ACD).
Design: Retrospective, noncomparative, interventional case series.
Methods: Medical charts of the patients with a low ACD of 2.8 mm or less who underwent implantable Collamer lens implantation were reviewed. To evaluate efficacy and safety, preoperative and postoperative uncorrected visual acuity, manifest refractive spherical equivalent (MRSE), endothelial cell count, intraocular pressure, angle opening distance at 500 μm from the scleral spur, trabecular-iris angle, and postoperative central vault were analyzed.
Results: The mean follow-up period for 18 eyes of 10 patients was 24.67 ± 15.09 months (range, 10 to 51 months). The preoperative ACD was 2.71 ± 0.08 mm (range, 2.42 to 2.79 mm). There was no postoperative endothelial cell count reduction to less than 2000 cells/mm(2) or intraocular pressure elevation to more than 21 mm Hg. The differences between preoperative and postoperative angle opening distance at 500 μm from the scleral spur and trabecular-iris angle were statistically significant (P = .000, both). The mean postoperative central vault was 331.83 ± 181.28 μm (range, 174 to 811 μm), which was significantly less than expected (P = .000). Anterior subcapsular cataract was observed in 2 eyes (11.1%).
Conclusions: Implantable Collamer lens implantation in shallow ACD eyes exhibited good visual outcomes. Severe complications did not develop over relatively long periods. Two cases of anterior subcapsular cataract developed in older aged patients, suggesting the need for thorough preoperative warning about the risk of cataract. In addition, a postoperative vault less than expected should be taken into account when determining implantable Collamer lens size in lower ACD eyes.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ajo.2014.06.005 | DOI Listing |
Int J Surg Case Rep
January 2025
King Khaled Eye Specialist Hospital, Riyadh, Kingdom of Saudi Arabia. Electronic address:
Introduction: Implantable collamer lens (ICL) is a posterior chamber phakic intraocular lens. It is usually indicated for high refractive error correction that cannot be treated by laser vision correction due to physical limitations.
Case Presentation: 39 years old male with a past ocular history of keratoconus underwent astigmatic keratotomy followed by crosslinking in both eyes a couple of years later due to signs of corneal ectasia progression.
Purpose: To evaluate dynamic changes in ciliary parameters and Implantable Collamer Lens V4C (ICL) (STAAR Surgical) haptic position using mydriatic and miotic agents and their effects on the central and peripheral vault.
Methods: This study involved 80 eyes from 40 consecutive patients (mean age: 28.05 years; range: 19 to 42 years) examined 3 months after ICL implantation.
Purpose: To evaluate the 12-month rotational stability and outcomes of V4c toric Implantable Collamer Lenses (TICLs) (STAAR Surgical) located at two different lens orientations (horizontal or oblique).
Methods: This retrospective study included a total of 403 eyes with TICL implantation, enrolled between January 2021 and December 2022, that were divided into two groups based on the minimum intended angle off the horizontal axis: horizontal with 263 eyes at an angle ±22.5 degrees or less and oblique with 140 eyes at a minimum intended angle of greater than ±22.
Sci Rep
January 2025
Department of Ophthalmology, Parkway Gleneagles Medical and Surgical Center, 389 Nanjing Xi Rd, Shanghai, 200032, China.
This study aims to compare the efficacy and safety of femtosecond laser-assisted implantable collamer lens (ICL) implantation with traditional manual ICL techniques. A retrospective analysis was conducted on patients who underwent ICL implantation at Beijing New Vision Eye Hospital in 2023. Patients (aged 18-45) were matched for gender and refractive error, with forty-two eyes in each group.
View Article and Find Full Text PDFGraefes Arch Clin Exp Ophthalmol
December 2024
Eye Ear Nose and Throat Hospital, Fudan University, XuHui District, No. 19 BaoQing Road, Shanghai, 200031, China.
Purpose: This study aimed to measure the changes in refraction difference value (RDV) in different parts of the retina before and after implantable collamer lenses (ICL) surgery.
Design: Prospective study.
Methods: This study included 40 eyes of 40 patients who underwent ICL implantation for myopia.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!