A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Pocket echocardiography system for detection of patent ductus arteriosus in neonates. | LitMetric

Pocket echocardiography system for detection of patent ductus arteriosus in neonates.

Echocardiography

Department of Pediatrics/Division of Neonatology, University of Florida, Gainesville, Florida.

Published: February 2015

Background: Neonates are commonly referred for a cardiology consult and an echocardiogram to rule out patent ductus arteriosus (PDA).

Objectives: Evaluate the usefulness of current pocket echocardiography system (PES) in PDA detection compared to traditional full-featured echo system (FFES).

Hypothesis: The determination of the presence of a PDA in neonates can be done using PES.

Methods: Fifty newborns with orders for echo evaluation were included in this study. A 5-minute PES scan was performed first. Then a full echo study was performed on a traditional FFES. Images were evaluated by three pediatric cardiologists blinded to the patients and the FFES results.

Results: The overall accuracy of reviewers rating PES versus FFES to rule in PDA had low false-positive rates 9.5% (95% CI: 1.2-30%), 11.8% (95% CI: 1.5-36%), 11.1% (95% CI: 1.4-35%) and the false-negative rate to rule out PDA was 0% (95% CI: 0-18%), 5.5% (95% CI: 0.14-27%), 0% (95% CI: 0-26%) for each reviewer 1, 2, and 3, respectively. Upon further analysis, PES scan was shown not to be accurate on infants with body weight below 1000 g with encouraging results for infants above 1000 g and those >37 weeks gestational age.

Conclusions: Our results suggest that the current PES could potentially be used in larger and near-term infants but has less than acceptable performance in low birth weight and premature infants in determining who should undergo a FFES study for persistent PDA diagnosis. Improvements in the technology along with developing limited training protocols for noncardiology personnel should make it possible for PES scan to be used as a screening tool and as an extension of physical examination especially in limited resource settings.

Download full-text PDF

Source
http://dx.doi.org/10.1111/echo.12656DOI Listing

Publication Analysis

Top Keywords

pes scan
12
pocket echocardiography
8
echocardiography system
8
patent ductus
8
ductus arteriosus
8
rule pda
8
95%
7
pes
6
pda
5
system detection
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!