Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objectives: To assess the inter-machine reliability of the Biodex System 3 Pro and Cybex Humac Norm Model 770 dynamometers for knee extensor and knee flexor peak torque measurements in isometric, concentric and eccentric tests.
Design: Randomized/crossover.
Setting: Exercise Research Laboratory, Federal University of Rio Grande do Sul (Brazil).
Participants: 25 healthy male subjects.
Main Outcome Measures: Isometric, concentric and eccentric knee extensor and knee flexor peak torques recorded in the same test procedure performed on both isokinetic dynamometers. One-way ANOVA, intraclass correlation coefficient (ICC), standard error of measurement (SEM) and coefficient of variation (CV) were used to verify significant differences, relative and absolute reliability between devices.
Results: No significant differences were found between tests performed on Biodex and Cybex (p > 0.05). ICC values indicated a high to very high reproducibility for isometric, concentric and eccentric peak torques (0.88-0.92), and moderate to high reliability for agonist-antagonist strength ratios (0.62-0.73). Peak torque did not show great difference between dynamometers for SEM (3.72-11.27 Nm) and CV (5.27-7.77%). Strength ratios presented CV values of 8.57-10.72%.
Conclusion: Maximal knee extensor and knee flexor tests performed in isometric (60° of knee flexion), concentric and eccentric modes at 60°/s in Biodex and Cybex dynamometers present similar values.