Purpose: To compare MRI sequences for breast density measurements on a 3T MRI system using IDEAL (Iterative Decomposition of water and fat with Echo Asymmetry and Least squares estimation) as possible physiology-like reference.

Materials And Methods: MRI examination was performed in 48 consecutive patients (mean age 41, years; range, 35-67 years) on a 3.0T scanner and 46 were included. All (fertile) women, were examined between days 5 and 15 of their menstrual cycle. MRI protocol included: T1-turbo spin-echo (T1-tSE), T2-turbo spin-echo (T2-tSE), VIBRANT (Volume Imaging for Breast Assessment) before and after injection of contrast media and IDEAL. Breast density was calculated with semi-automated software. Statistical analysis was performed with non-parametric tests.

Results: Mean percentage of breast density calculated in each sequence was: T1-tSE = 56%; T2-tSE  = 52%; IDEAL FatOnly  = 55%; IDEAL WaterOnly  = 53%, VIBRANT = 55%. Significant differences were observed between T2-tSE and both T1-tSE (p<0.001), VIBRANT sequences (p = 0.009), T1-tSE and both IDEAL WaterOnly (p = 0.007) and IDEAL FatOnly (p = 0.047). Breast density percentage showed a positive linear correlation among different sequences: r ≥ 0.93.

Conclusions: Differences exist between MRI sequences used to assess breast density percentage. T1-weighted sequences values were similar to IDEAL sequences.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4044003PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0099027PLOS

Publication Analysis

Top Keywords

breast density
16
mri system
8
density calculated
8
breast
5
mri
5
density assessment
4
assessment mri
4
system comparison
4
comparison sequences
4
sequences purpose
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!