Comparison of a conventional HIV 1/2 line immunoassay with a rapid confirmatory HIV 1/2 assay.

J Virol Methods

Blood Transfusion Service SRC Berne, Murtenstrasse 133, 3008 Berne, Switzerland. Electronic address:

Published: September 2014

The performance of the rapid confirmatory HIV 1/2 assay Geenius was compared with the conventional HIV 1/2 line immunoblot (INNO-LIA HIV I/II Score). One hundred HIV 1/2 confirmed positive samples from donors and patients and 136 negative screening samples from blood donors were evaluated with both assays. A 20 member performance panel consisting of different HIV 1 and 2 subtypes was also analysed. Ninety-nine of the confirmed HIV positive samples were positive with both assays. One sample was positive with the INNO-LIA HIV I/II Score but indeterminate with the Geenius HIV 1/2. From 136 negative blood donor samples (negative with a combo HIV assay and a highly sensitive ID-NAT), 125 were concordant negative. Six and five samples were incorrectly indeterminate with the INNO-LIA HIV I/II Score and the Geenius HIV 1/2, respectively. One sample was weak positive with the INNO-LIA HIV I/II Score but negative with the Geenius HIV 1/2. The 20 member performance showed equivalent results with both assays. The rapid assay showed a comparable sensitivity and specificity for confirmation for positive and negative HIV donor and patient samples as well for a 20 member performance panel.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jviromet.2014.05.010DOI Listing

Publication Analysis

Top Keywords

hiv 1/2
32
hiv
16
inno-lia hiv
16
hiv i/ii
16
i/ii score
16
member performance
12
geenius hiv
12
conventional hiv
8
1/2
8
rapid confirmatory
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!