Dosimetric adaptive IMRT driven by fiducial points.

Med Phys

Department of Oncology, Laboratory of Experimental Radiotherapy, KU Leuven, Herestraat 49, 3000 Leuven, Belgium and Department of Oncology, MRC-CR-UK Gray Institute of Radiation Oncology and Biology, University of Oxford, Oxford OX1 2JD, United Kingdom.

Published: June 2014

Purpose: Intensity modulated radiotherapy (IMRT) and volumetric modulated arc therapy have become standard treatments but are more sensitive to anatomical variations than 3D conformal techniques. To correct for inter- and intrafraction anatomical variations, fast and easy to implement methods are needed. Here, the authors propose a full dosimetric IMRT correction that finds a compromise in-between basic repositioning (the current clinical practice) and full replanning. It simplifies replanning by avoiding a recontouring step and a full dose calculation. It surpasses repositioning by updating the preoptimized fluence and monitor units (MU) using a limited number of fiducial points and a pretreatment (CB)CT. To adapt the fluence the fiducial points were projected in the beam's eye view (BEV). To adapt the MUs, point dose calculation towards the same fiducial points were performed. The proposed method is intrinsically fast and robust, and simple to understand for operators, because of the use of only four fiducial points and the beam data based point dose calculations.

Methods: To perform our dosimetric adaptation, two fluence corrections in the BEV are combined with two MU correction steps along the beam's path. (1) A transformation of the fluence map such that it is realigned with the current target geometry. (2) A correction for an unintended scaling of the penumbra margin when the treatment beams scale to the current target size. (3) A correction for the target depth relative to the body contour and (4) a correction for the target distance to the source. The impact of the correction strategy and its individual components was evaluated by simulations on a virtual prostate phantom. This heterogeneous reference phantom was systematically subjected to population based prostate transformations to simulate interfraction variations. Additionally, a patient example illustrated the clinical practice. The correction strategy was evaluated using both dosimetric (CTV mean dose, conformity index) and clinical (tumor control probability, and normal tissue complication probability) measures.

Results: Based on the current experiments, the intended target dose and tumor control probability could be assured by the proposed method (TCP ≥ TCP(intended)). Additionally, the conformity index error was more than halved compared to the current clinical practice (ΔCI(95%) from 40% to 16%) resulting in improved organ at risk protection. All the individual correction steps had an added value to the full correction.

Conclusions: A limited number of fiducial points (no organ contours required) and an in-room (CB)CT are sufficient to perform a full dosimetric correction for IMRT plans. In the presence of interfraction variation, the corrected plans show superior dose distributions compared to our current clinical practice.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4876378DOI Listing

Publication Analysis

Top Keywords

fiducial points
24
clinical practice
16
current clinical
12
correction
9
anatomical variations
8
full dosimetric
8
dose calculation
8
limited number
8
number fiducial
8
point dose
8

Similar Publications

This paper deals with neuro-registration using tele-manipulation (Master-Slave Manipulation) to facilitate tele-surgery and enhance the overall accuracy and reach of the robot-assisted neurosurgery. Accurate Neuro-registration is important as the success of the surgical procedure highly depends on it. A 6-degree-of-freedom Parallel Kinematic Mechanism (6D-PKM) master-slave robot in tele-manipulation mode is utilized for both neuro-registration and neurosurgery.

View Article and Find Full Text PDF

Traditional marker-based photogrammetry systems often require the attachment and removal of a sticker for each measured point, involving labor-intensive manual steps. This paper presents an innovative approach that utilizes raised, cross-shaped markers, referred to as 'molded markers', directly embedded into composite pieces. In this study, these markers, commonly employed in other industrial processes, serve as fiducial markers for accurate photogrammetry.

View Article and Find Full Text PDF

Introduction: Recognizing human actions is crucial for allowing machines to understand and recognize human behavior, with applications spanning video based surveillance systems, human-robot collaboration, sports analysis systems, and entertainment. The immense diversity in human movement and appearance poses a significant challenge in this field, especially when dealing with drone-recorded (RGB) videos. Factors such as dynamic backgrounds, motion blur, occlusions, varying video capture angles, and exposure issues greatly complicate recognition tasks.

View Article and Find Full Text PDF

Background: Cable-driven continuum manipulators (CDCMs) enable scar-free procedures but face limitations in workspace and control accuracy due to hysteresis.

Methods: We introduce an extensible CDCM with a semi-active mechanism (SAM) and develop a real-time hysteresis compensation control algorithm using a temporal convolution network (TCN) based on data collected from fiducial markers and RGBD sensing.

Results: Performance validation shows the proposed controller significantly reduces hysteresis by up to 69.

View Article and Find Full Text PDF

Distributed minimum error entropy with fiducial points Kalman filter for state tracking.

ISA Trans

January 2025

Key Laboratory of Magnetic Suspension Technology and Maglev Vehicle, Ministry of Education, School of Electrical Engineering, Southwest Jiaotong University, Chengdu, 610031, China. Electronic address:

With the growing size of the system, this distributed Kalman filter (DKF) is widely used in multi-sensor networks. However, it is difficult for DKF to accurately estimate state values in non-Gaussian noise environments. In this paper, a regression equation is first constructed to contain all sensor node information.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!