Background: Continuous glucose monitoring using subcutaneously inserted sensors currently requires blood glucose tests for sensor calibration. Alternatively, sensors precalibrated during the manufacturing process may eliminate the need for fingerstick calibrations. In this study we evaluated the feasibility of sensor factory calibration in subjects with diabetes.

Methods: A total of 33 subjects with diabetes were asked to wear 4 sensors in parallel, 2 on the arm and 2 on the abdomen. Sensors from a lot with low in vitro sensitivity coefficient of variation were used in the study. Based on frequent capillary blood glucose measurements, the average glucose sensitivity of each sensor was determined over a 5-day wear time. The in vivo sensitivities were analyzed for inter- and intrasubject variation. Mean absolute relative difference (MARD) calculation and consensus error grid analysis (EGA) were performed using a single calibration factor for all sensors, to simulate factory calibration and compared against conventional finger-stick calibration.

Results: The sensitivity coefficient of variation between sensors increased from 2.9% in vitro to 6.0% in vivo. No difference in sensor response between subjects (P = .069) as well as between insertion sites (arm and abdomen) was detected (P = .104). Applying one calibration factor to all sensors in the study resulted in an MARD of 13.4%, and 83.5% of the values fell in consensus EGA zone A. Multiple fingerstick calibration resulted in an MARD of 12.7% and 84.1% in zone A.

Conclusions: Feasibility of factory calibration was demonstrated in subjects with diabetes using sensors based on "wired enzyme" technology, resulting in accuracy metrics similar to sensors calibrated with capillary blood glucose.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4454101PMC
http://dx.doi.org/10.1177/1932296813511747DOI Listing

Publication Analysis

Top Keywords

factory calibration
16
subjects diabetes
12
blood glucose
12
sensors
10
feasibility factory
8
calibration
8
arm abdomen
8
sensitivity coefficient
8
coefficient variation
8
capillary blood
8

Similar Publications

Optimal Timing and Approaches for First Tangential Excision in Patients with ≥70% TBSA Burns: Insights from a Retrospective Cohort of a Mass Casualty Incident.

Ann Surg

January 2025

Department of Burn Surgery, the First Affiliated Hospital of Naval Medical University, 200433, Shanghai, People's Republic of China; Research Unit of Key Techniques for Treatment of Burns and Combined Burns and Trauma Injury, Chinese Academy of Medical Sciences, 200433, Shanghai, People's Republic of China.

Objective: We aim to determine the optimal timing and approaches for first tangential excision of severely burned patients ≥70% total body surface area (TBSA).

Background: Early tangential excision is the gold-standard surgical therapy for full-thickness burns. However, there are debates about its optimal timing and approaches for severely burned patients ≥70%TBSA.

View Article and Find Full Text PDF

Quantifying the Impact of Soil Moisture Sensor Measurements in Determining Green Stormwater Infrastructure Performance.

Sensors (Basel)

December 2024

Department of Civil and Environmental Engineering, Villanova University, 800 Lancaster Avenue, Villanova, PA 19085, USA.

The ability to track moisture content using soil moisture sensors in green stormwater infrastructure (GSI) systems allows us to understand the system's water management capacity and recovery. Soil moisture sensors have been used to quantify infiltration and evapotranspiration in GSI practices both preceding, during, and following storm events. Although useful, soil-specific calibration is often needed for soil moisture sensors, as small measurement variations can result in misinterpretation of the water budget and associated GSI performance.

View Article and Find Full Text PDF

Home Urine Dipstick Screening for Bladder and Kidney Cancer in High-Risk Populations in England: A Microsimulation Study of Long-Term Impact and Cost-Effectiveness.

Pharmacoeconomics

January 2025

Sheffield Centre for Health and Related Research (SCHARR), School of Medicine and Population Health, The University of Sheffield, Regent Court, 30 Regent Street, Sheffield, UK.

Background: Testing high-risk populations for non-visible haematuria may enable earlier detection of bladder cancer, potentially decreasing mortality. This research aimed to assess the cost-effectiveness of urine dipstick screening for bladder cancer in high-risk populations in England.

Methods:  A microsimulation model developed in R software was calibrated to national incidence data by age, sex and stage, and validated against mortality data.

View Article and Find Full Text PDF

The problem of calibrating phased array antennas in a noisy environment using an autocorrelation algorithm is investigated and a mathematical model of the autocorrelation calibration method is presented. The proposed calibration system is based on far-field scanning of the phased array antenna in an environment with internal noise and external interference. The proposed method is applied to a phased array antenna and compared with traditional rotating-element electric-field vector methods, which involve identifying the maximum and minimum vector-sum points (REVmax and REVmin, respectively).

View Article and Find Full Text PDF

Evaluation of Different Commercial Sensors for the Development of Their Automatic Irrigation System.

Sensors (Basel)

November 2024

Centro de Investigaciones Científicas y Tecnológicas de Extremadura (CICYTEX), Finca La Orden, Regional Government of Extremadura, Highway A-V, Km 372, 06187 Guadajira, Badajoz, Spain.

Reliable soil moisture information is essential for accurate irrigation scheduling. A wide range of soil moisture sensors are currently available on the market, but their performance needs to be evaluated as most sensors are calibrated under limited laboratory conditions. The aim of this study was to evaluate the performance of six commercially available moisture sensors (HydraProbe, Teros 10, Teros 11, EnviroPro, CS616 and Drill & Drop) and three tensiometers (Irrometer RSU-C-34, Teros 32 and Teros 21) and to establish calibration equations for a typical sandy soil of the Doñana National Park (Huelva, Spain).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!