Delayed fracture of MatrixRIB precontoured plate system.

Interact Cardiovasc Thorac Surg

Department of Surgery, Prince of Wales Hospital, The Chinese University of Hong Kong, Hong Kong SAR, China.

Published: September 2014

Objectives: Use of titanium prostheses for reconstructing chest wall defects following major chest wall resections is rapidly increasing in popularity. Previously, complications including prosthesis fracture have been reported for the system secured to the rib ends using clips following chest wall reconstruction and pectus excavatum repair. By contrast, fracture failure in titanium systems fastening the plate to the rib with locking screws through predrilled holes has not been previously reported, possibly owing to differences in the design and material of the system.

Methods: We report an index case of plate fracture in the latter design following rib reconstruction and discuss the pathophysiology and relative risks behind such fractures.

Results: Fracture of titanium plate occurred after chest impact during sport at 25 months following initial plate implantation. Surgical excision was performed because of persistent pain and cough.

Conclusions: Fracture failure can occur in those systems fastening the plate to the rib with locking-screws, and patients, particularly those engaged in contact sports or occupations with chest impact risks, should be informed of this possibility during consent, as surgical removal is most likely required following implant fracture failur.

Download full-text PDF

Source
http://dx.doi.org/10.1093/icvts/ivu175DOI Listing

Publication Analysis

Top Keywords

chest wall
12
fracture failure
8
systems fastening
8
fastening plate
8
plate rib
8
chest impact
8
plate
6
fracture
6
chest
5
delayed fracture
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!