Introduction: The purpose of this in vitro study was to obtain a better understanding of the mechanism of irrigant traveling apically and generating 2-phase intermittent counter flow in straight root canal models during activation of the irrigant by ultrasonic means in an endodontic procedure.
Methods: A high-speed imaging system, with high temporal and spatial resolution (FastCam SA5; Photron, Tokyo, Japan) at a frame rate of 100,000 frames per second using a macro lens (60 mm, f/2.8; Nikon, Tokyo, Japan), was used to visualize, in glass models of root canals, an ultrasonically induced acoustic pressure wave in an EDTA solution environment. A 25-mm stainless steel noncutting file #20 driven by an ultrasonic device (P5 Newtron; Satelec Acteon, Mérignac, France) at power settings of 5 and 7 produced disturbances at the solution-air interface.
Results: We found that apically directed travel of the irrigant was caused by disruption of the surface tension at the solution-air interface. This disruption caused by ultrasonic activation energy displaced air in the form of bubbles from the apical region toward the solution.
Conclusions: The apical movement of the solution may be attributed to ultrasonically induced wave generation at the solution-air interface, resulting in the removal of trapped air from the root canal and allowing the solution to travel apically in the opposite directions (via a 2-phase intermittent counter flow).