Purpose: We developed a stereotactic device to guide the puncture for percutaneous nephrolithotripsy, which uses the distance from the target calyx to its perpendicular point on skin (SCD) to calculate the needle´s entry angle. This study seeks to validate the use of measurements obtained by ultrasound (US) and computerized tomography (CT) for needle´s entry angle calculation and to study factors that may interfere in this procedure.
Materials And Methods: Height, weight, abdominal circumference, CT of the urinary tract in dorsal decubitus (DD) and ventral decubitus (VD), and US of the kidneys in VD were obtained from thirty-five renal calculi patients. SCD obtained were compared and correlated with body-mass index (BMI).
Results: BMI was 28.66 ± 4.6 Kg/m2. SCD on CT in DD was 8.40 ± 2.06cm, in VD was 8.32 ± 1.95cm, in US was 6.74 ± 1.68cm. SCD measured by US and CT were statistically different (p < 0.001), whereas between CT in DD and VD were not. SCD of the lower calyx presented moderate correlation with BMI.
Conclusion: SCD obtained by CT in ventral and dorsal decubitus may be used for calculation of the needle´s entry angle. SCD obtained by US cannot be used. A rule for the correlation between BMI and the SCD could not be determined.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1590/S1677-5538.IBJU.2014.02.11 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!