Cardiac resynchronization therapy (CRT) emerged 2 decades ago as a useful form of device therapy for heart failure associated with abnormal ventricular conduction, indicated by a wide QRS complex. In this Review, we present insights into how to achieve the greatest benefits with this pacemaker therapy. Outcomes from CRT can be improved by appropriate patient selection, careful positioning of right and left ventricular pacing electrodes, and optimal timing of electrode stimulation. Left bundle branch block (LBBB), which can be detected on an electrocardiogram, is the predominant substrate for CRT, and patients with this conduction abnormality yield the most benefit. However, other features, such as QRS morphology, mechanical dyssynchrony, myocardial scarring, and the aetiology of heart failure, might also determine the benefit of CRT. No single left ventricular pacing site suits all patients, but a late-activated site, during either the intrinsic LBBB rhythm or right ventricular pacing, should be selected. Positioning the lead inside a scarred region substantially impairs outcomes. Optimization of stimulation intervals improves cardiac pump function in the short term, but CRT procedures must become easier and more reliable, perhaps with the use of electrocardiographic measures, to improve long-term outcomes.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1038/nrcardio.2014.67 | DOI Listing |
Eur Heart J Case Rep
January 2025
Department of Cardiology, Yokohama City University Hospital, 3-9 Fukuura, Kanazawa-Ku, Yokohama city, Kanagawa 236-0004, Japan.
Background: Coronary sinus (CS) lead placement in persistent left superior vena cava (PLSVC) cases is challenging because of the poor backup force of the guiding catheter within the enlarged CS. Active fixation Quadripolar leads (Attain Stability™ Quad 4798, Medtronic) can expand choice to CS branches with limited access; however, no cases of anchoring to the main body of the CS have been published to date.
Case Summary: We describe a case of cardiac resynchronization therapy pacemaker upgrade in a 79-year-old female who developed pacing-induced cardiomyopathy after pacemaker implantation via the right superior vena cava (SVC) for atrioventricular block eight years ago wherein PLSVC was revealed during the procedure.
Cardiol Young
January 2025
Department of Medical Education and Clinical Sciences, Elson S Floyd College of Medicine, Washington State University, Spokane, WA, USA.
Sudden cardiac death poses a significant risk in patients with surgically repaired tetralogy of Fallot. Despite extensive research, risk stratification practices vary. This study surveyed the Pediatric and Adult Congenital Electrophysiology Society to identify these differences.
View Article and Find Full Text PDFPacing Clin Electrophysiol
January 2025
Department of Cardiology, General Hospital of Northern Theater Command, Shenyang, People's Republic of China.
Background: The evidence in the primary prevention of sudden cardiac death (SCD) by using implantable cardioverter defibrillators (ICD) in patients with ischemic cardiomyopathy (ICM) is well-established but remains controversial for those with nonischemic cardiomyopathy (NICM). This study evaluates whether prognostic differences exist between ICM and NICM patients after ICD implantation.
Methods: Patients initially undergoing ICD implantation as primary SCD prevention were retrospectively enrolled from January 2017 to May 2023.
Pacing Clin Electrophysiol
January 2025
Aurora Cardiovascular and Thoracic Services, Aurora Sinai/Aurora St. Luke's Medical Centers, Milwaukee, Wisconsin, USA.
Background: Concurrent Micra leadless pacemaker (Medtronic, Minneapolis, Minnesota) implantation and atrioventricular node (AVN) ablation has been shown to be feasible and safe in patients with symptomatic, drug-refractory atrial fibrillation (AF). However, major complications within the 30 days after concurrent Micra implantation and AVN ablation have been reported. We evaluated the efficacy and safety of the concurrent procedure at our institution.
View Article and Find Full Text PDFPacing Clin Electrophysiol
January 2025
Second Division of Cardiology, Cardiac-Thoracic and Vascular Department, University Hospital of Pisa, Pisa, Italy.
This case details the successful implantation of a leadless pacemaker following the extraction of transvenous leads in a 72-year-old female patient with a complex cardiovascular history. The patient had undergone a series of cardiac interventions, including a recent percutaneous tricuspid valve repair with a metal clip implant due to severe regurgitation. After presenting with an infection at the pacemaker site, methicillin-resistant Staphylococcus hominis was identified, necessitating the removal of the entire pacing system.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!