The new mid-scala electrode array: a radiologic and histologic study in human temporal bones.

Otol Neurotol

*Department of Otorhinolaryngology-Head and Neck Surgery, †Department of Radiology, and ‡Institute of Musicians' Medicine, Freiburg University Medical Center, Freiburg, Germany.

Published: September 2014

Hypothesis: To analyze the quality of insertion of the newly developed midscala (MS) electrode, which targets a midscalar electrode position to reduce the risk of trauma to the lateral wall and the modiolus.

Background: Modern cochlear implant surgery aims for a safe intracochlear placement of electrode arrays with an ongoing debate regarding cochleostomy or round window (RW) insertion and the use of lateral wall or perimodiolar electrode placement. Intracochlear trauma after insertion of different electrodes depends on insertion mode and electrode design and may result in trauma to the delicate structures of the cochlear.

Methods: We performed a temporal bone (TB) trial with insertion of the MS electrode in n = 20 TB's after a mastoidectomy and posterior tympanotomy. Insertion was performed either via the RW or a cochleostomy. Electrode positioning, length of insertion, and angle of insertion were analyzed with rotational tomography (RT). TBs were histologically analyzed. Results of RT and histology were compared.

Results: Scala tympani (ST) insertion could be accomplished reliably by both RW and via a cochleostomy approach. In 20 TBs, 1 scala vestibuli insertion, 1 incomplete (ST), and 1 elevation of basilar membrane were depicted. No trauma was found in 94.7% of all ST insertions. RT allowed determination of the intracochlear electrode position, which was specified by histologic sectioning.

Conclusion: The new MS electrode seems to fulfill reliable atraumatic intracochlear placement via RW and cochleostomy approaches. RT is available for evaluation of intracochlear electrode position, serving as a potential quality control instrument in human implantation.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAO.0000000000000412DOI Listing

Publication Analysis

Top Keywords

electrode position
12
insertion
10
electrode
10
lateral wall
8
intracochlear placement
8
intracochlear electrode
8
intracochlear
5
mid-scala electrode
4
electrode array
4
array radiologic
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!