Objective: To measure the orthodontic forces exerted on the upper right central incisor with the increase of the distance of tooth movement and the thickness of the aligner.

Methods: The labial movement of upper right central incisor at various distances (0.3, 0.6, 0.9, 1.2, 1.5, 1.8 mm) was designed and the stereolithography model (2 times bigger than the original model) was created with 3-D scanning and tomography output. These models were used to fabricate the aligners with different thicknesses (0.8, 1.0, 1.5 and 2.0 mm) of the thermoplastic materials (6 samples for various distances of tooth movement and thicknesses). Orthodontic forces exerted on the upper right central incisor were measured with the micro-stress sensor measurement system.

Results: The orthodontic forces increased with the increase of the thickness of the aligner at the same distance of tooth movement (P < 0.05). The orthodontic force was (1.237 ± 0.082), (1.543 ± 0.059), (3.602 ± 0.102), (6.734 ± 0.063) N when the labial movement of upper right central incisor was 0.3 mm with the aligner of 0.8, 1.0, 1.5, 2.0 mm. The orthodontic forces also increased with the increase of the distance of the tooth movement at the same thickness of the aligner(P < 0.05). The orthodontic force of the aligner of 0.8 mm were (1.354 ± 0.039), (1.288 ± 0.037), (1.479 ± 0.031), (1.799 ± 0.039) N when the upper right central tooth labial movement at 0.6, 0.9, 1.2, 1.5 mm.

Conclusions: The orthodontic forces increased with the increase of the distance of tooth movement and the thickness of the aligner.

Download full-text PDF

Source

Publication Analysis

Top Keywords

orthodontic forces
24
upper central
24
tooth movement
24
central incisor
20
distance tooth
20
increase distance
16
movement thickness
16
forces exerted
12
exerted upper
12
labial movement
12

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!