BIOLOX delta is the newest ceramic composite material of the BIOLOX family and its use in hip devices represents a good alternative to Metal-on-Metal or Metal-on-Polyethylene coupling. Some Metal-on-Metal hip devices have shown chromium and cobalt release, which promoted toxic effects. Furthermore, the ceramic composite contains trivalent chromium, even if strongly bound to the alumina lattice. The present study is aimed at detecting any 'in vivo' release of chromium ions from BIOLOX delta bearings in the blood, erythrocytes and urine of patients. Twenty patients implanted with total hip arthroplasty (THA) with BIOLOX delta-BIOLOX delta couplings and 21 subjects with no implanted prostheses were studied. Inductively coupled plasma mass spectrometry equipped with dynamic reaction cell was used for analysis. In the THA group the Cr ions values were; in blood mean 0.21 µg/l (±0.09), in serum 0.21 µg/l (±0.12), in normalized erythrocytes 0.13 µg/l (±0.09), in normalized urine 0.12 µg/g creatinine (±0.13). In the control group the Cr ions values were; in blood mean 0.22 µg/l (st dev 0.14), in serum 0.17 µg/l (±0.08), in normalized erythrocytes 0.13 µg/l (±0.11), in normalized urine 0.07 µg/g creatinine (±0.08). The Lab reference values were 0.1-5.0 µg/l for blood, 0.1-0.5 µg/l for serum, 0.14-4.58 µg/l for normalised erythrocytes and 0.05-2.2 µg/l for urine. All samples in both groups resulted in chromium levels within the normal reference range and the safety of BIOLOX delta ceramics, in terms of chromium ions release, has been demonstrated.

Download full-text PDF

Source
http://dx.doi.org/10.5301/hipint.5000132DOI Listing

Publication Analysis

Top Keywords

biolox delta
12
µg/l
10
ceramic composite
8
hip devices
8
chromium ions
8
group ions
8
ions values
8
values blood
8
021 µg/l
8
µg/l ±009
8

Similar Publications

Atraumatic Fracture of Newer Generation Ceramic Head Three Days Post-op: A Case Report.

Cureus

December 2024

Orthopedic Surgery, Tri-County Orthopedics, Bridgewater, USA.

A 73-year-old female experienced an atraumatic fracture of a BIOLOX delta ceramic femoral head following uncomplicated right total hip arthroplasty using a ceramic-on-polyethylene bearing. The fracture occurred post-operatively, as revealed by radiography after the patient reported a clunking sensation and leg shortening. Revision surgery involved replacing the fractured head and liner with careful removal of ceramic debris.

View Article and Find Full Text PDF

Background: Few studies compare complications and results in ceramic-ceramic (CeCe) metal-on-crosslink polyethylene (Me-XLPE) hip prosthesis.

Methods: of 2 series of hydroxyapatite-coated Furlong Active implanted with identical protocol by 1 surgeon in 1 hospital from 2006 to 2014.

Bearing Surface: CeCe (Biolox Forte/Delta) in 35 cases of 53.

View Article and Find Full Text PDF

(1) Background: The kinematic characteristics of disc prosthesis undergoing complex motion are not well understood. Therefore, examining complex motion may provide an improved understanding of the post-operative behavior of spinal implants. (2) Methods: The aim of this study was to develop kinematic tests that simulate multiplanar motion and combined rotational-translational motion in a disc prosthesis.

View Article and Find Full Text PDF

Introduction: A fracture of the ceramic head in bipolar hemiarthroplasty using an inner polyliner has not been reported yet, and there seems to be no report of simultaneous breakage of the fourth-generation BIOLOX Delta ceramic head and liner in total hip arthroplasty.

Method: A 44-year-old male patient underwent bipolar hemiarthroplasty using a third-generation BIOLOX Forte ceramic head 3 years and 9 months earlier for osteonecrosis of femoral head (ONFH) and visited our hospital due to a ceramic head fracture. Conversion total hip arthroplasty was performed.

View Article and Find Full Text PDF

Material-Intrinsic NIR-Fluorescence Enables Image-Guided Surgery for Ceramic Fracture Removal.

Adv Healthc Mater

April 2024

Nanoparticle Systems Engineering Laboratory, Institute of Energy and Process Engineering (IEPE), Department of Mechanical and Process Engineering (D-MAVT), ETH Zurich, Sonneggstrasse 3, Zurich, 8092, Switzerland.

Hip arthroplasty effectively treats advanced osteoarthritis and is therefore entitled as "operation of the 20th century." With demographic shifts, the USA alone is projected to perform up to 850 000 arthroplasties annually by 2030. Many implants now feature a ceramic head, valued for strength and wear resistance.

View Article and Find Full Text PDF