Aim: Providing consistent levels of oxygen saturation (SpO2 ) for infants in neonatal intensive care units is not easy. This study explored how effectively the Auto-Mixer(®) algorithm automatically adjusted fraction of inspired oxygen (FiO2 ) levels to maintain SpO2 within an intended range in extremely low birth weight infants receiving supplemental oxygen without mechanical ventilation.

Methods: Twenty extremely low birth weight infants were randomly assigned to the Auto-Mixer(®) group or the manual intervention group and studied for 12 h. The SpO2 target was 85-93%, and the outcomes were the percentage of time SpO2 was within target, SpO2 variability, SpO2 >95%, oxygen received and manual interventions.

Results: The percentage of time within intended SpO2 was 58 ± 4% in the Auto-Mixer(®) group and 33.7 ± 4.7% in the manual group, SpO2 >95% was 26.5% vs 54.8%, average SpO2 and FiO2 were 89.8% vs 92.2% and 37% vs 44.1%, and manual interventions were 0 vs 80 (p < 0.05). Brief periods of SpO2  < 85% occurred more frequently in the Auto-Mixer(®) group.

Conclusion: The Auto-Mixer(®) effectively increased the percentage of time that SpO2 was within the intended target range and decreased the time with high SpO2 in spontaneously breathing extremely low birth weight infants receiving supplemental oxygen.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4228757PMC
http://dx.doi.org/10.1111/apa.12684DOI Listing

Publication Analysis

Top Keywords

spo2
9
receiving supplemental
8
supplemental oxygen
8
oxygen mechanical
8
extremely low
8
low birth
8
birth weight
8
weight infants
8
auto-mixer® group
8
spo2 target
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!