Estimating energy expenditure using heat flux measured at a single body site.

Med Sci Sports Exerc

1Division of Endocrinology, Metabolism, and Diabetes, University of Colorado Anschutz Medical Campus, Aurora, CO; 2Division of Geriatric Medicine, University of Colorado Anschutz Medical Campus, Aurora, CO; 3School of Public Health, University of Colorado Anschutz Medical Campus, Aurora, CO; and 4Necessity Consulting, Mars, PA.

Published: November 2014

Introduction: The Personal Calorie Monitor (PCM) is a portable direct calorimeter that estimates energy expenditure (EE) from measured heat flux (i.e., the sum of conductive, convective, radiative, and evaporative heat).

Purpose: The primary aim of this study was to compare EE estimated from measures of heat flux with those measured using indirect calorimetry in a thermoneutral environment (26°C). A secondary aim was to determine whether exposure to ambient temperature below thermoneutral condition (19°C) influences the accuracy of the PCM.

Methods: Thirty-four adults (mean ± SD: age, 28 ± 5 yr; body mass index, 22.9 ± 2.6 kg · m(-2)) were studied for 5 h in a whole-room indirect calorimeter (IC) in thermoneutral and cool conditions. Participants wore the PCM on their upper arm and completed two 20-min treadmill walking bouts (0% grade, 3 mph). The remaining time was spent sedentary (e.g., watching television, using a computer).

Results: In thermoneutral conditions, EE values (mean (95% confidence interval)) measured by IC and PCM were 560.0 (526.5-593.5) and 623.3 (535.5-711.1) kcal, respectively. In cool conditions, EE values measured by IC and PCM were 572.5 (540.9-604.0) and 745.5 (668.1-822.8) kcal, respectively. Under thermoneutral conditions, mean PCM minute-by-minute EE tracked closely with IC, resulting in a small nonsignificant bias (63 kcal (-5.8 to 132.4)). During cool conditions, mean PCM minute-by-minute EE did not track IC, resulting in a large bias (173.0 kcal (93.9-252.1)) (P <; 0.001).

Conclusions: This study demonstrated the validity of using measured heat flux to estimate EE. However, accuracy may be impaired in cool conditions possibly because of excess heat loss from the exposed limbs.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4199907PMC
http://dx.doi.org/10.1249/MSS.0000000000000346DOI Listing

Publication Analysis

Top Keywords

heat flux
12
cool conditions
12
energy expenditure
8
flux measured
8
thermoneutral conditions
8
conditions values
8
measured pcm
8
conditions pcm
8
pcm minute-by-minute
8
pcm
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!