Recent advances in and limitations of cardiac output monitoring by means of electrical impedance tomography.

Anesth Analg

From the Institute for Medical Information Technology, RWTH Aachen University, Aachen, Germany; Department of Anesthesia and Intensive Care, Sahlgrenska University Hospital Gothenburg, Gothenburg, Sweden; and Department of Medical Statistics, RWTH Aachen University, Aachen, Germany.

Published: July 2014

Background: Currently, the monitoring of cardiac output (CO) and stroke volume (SV) is mainly performed using invasive techniques. Therefore, performing CO monitoring noninvasively by means of electrical impedance tomography (EIT) would be advantageous for intensive care. Our hypothesis was that, by means of EIT, it is possible to assess heart rate (HR) and to quantify changes in SV due to changes in ventilator settings.

Methods: CO (HR and SV) of 14 pigs (32-40 kg body weight) was changed by incremental increases in positive end-expiratory pressure levels (0, 5, 10, 15, and 20 cm·H2O; ramp maneuver). This ramp maneuver was applied 4 times in each animal, yielding 43 evaluable single experiments. At each positive end-expiratory pressure level, SV was assessed by transpulmonary thermodilution using a PiCCO device. EIT data were acquired using a Dräger EIT Evaluation Kit 2.

Results: The EIT-based SV-related signal, Z(SV) (in [AU]), showed only a weak correlation (after excluding 2 measurements) with SV(TTD) of r = 0.58 (95% confidence interval, 0.43-0.71). If Z(SV) is calibrated by the reference 1 time for each experiment (defined as SVEIT), the correlation is approximately 0.85 (95% confidence interval, 0.78-0.90). A possible reason for the moderate correlation is the unexpected scaling pattern, leading to amplification of the cardiac impedance signal, found in some animals. The scaling is probably due to the imperfect reconstruction (i.e., a change of sensitivity) of the EIT images or to a change in the position of the heart.

Conclusions: The hypothesis that EIT can be used to monitor CO and SV was confirmed, but further studies are required before this technique can be applied in clinical practice. HR was determined robustly and accurately. For SV monitoring, promising results were obtained in 80% of the experiments. However, unexpected scaling of the cardiac EIT signal causing inaccurate estimation of SV remains an issue. Before robust assessment of SV by EIT is suitable for clinical practice, the cause of and compensation for undesired scaling effects need to be investigated.

Download full-text PDF

Source
http://dx.doi.org/10.1213/ANE.0000000000000241DOI Listing

Publication Analysis

Top Keywords

cardiac output
8
electrical impedance
8
impedance tomography
8
eit
8
hypothesis eit
8
positive end-expiratory
8
end-expiratory pressure
8
ramp maneuver
8
95% confidence
8
confidence interval
8

Similar Publications

Purpose: This study explores the effects of anthracycline chemotherapy (AC) on breast cancer patients, focusing on changes in body composition, advanced echocardiographic parameters at rest and during exercise, and biomarkers; and subsequently assesses whether these parameters are associated with impaired cardiorespiratory fitness (CRF).

Methods: In this prospective study, we evaluated women with early-stage breast cancer undergoing AC at three visits: before AC, 1 month after, and 6 months post-AC.

Results: The study included 32 women with breast cancer, with functional disability increasing from 9.

View Article and Find Full Text PDF

Background: The role of left ventricular global longitudinal strain (LVGLS) in coronary artery bypass grafting (CABG) and outcomes such as low cardiac output syndrome (LCOS) is not well established. The authors investigated the relationship between LVGLS before and after induction of anesthesia, their differences, and their relationship with LCOS and other outcomes.

Methodology: A prospective observational study was conducted in a public/private hospital with 50 adult patients scheduled for on-pump CABG with normal left ventricular ejection fraction (LVEF).

View Article and Find Full Text PDF

Purpose: Catheter ablation (CA) for atrial fibrillation (AF) in heart failure patients with preserved ejection fraction (HFPEF) has shown promising results in reducing mortality and improving heart function. However, previous studies have been limited by a lack of control groups and significant heterogeneity in their methodologies.

Hypothesis: CA for AF in HFPEF patients may not increase the complications and had similarly the rate of freedom from AF vs.

View Article and Find Full Text PDF

Background: ST-elevation myocardial infarction (STEMI) demands near-time reperfusion to reduce the risk of long-term heart failure. This study evaluates the proportion of impaired left ventricular ejection fraction (LVEF) following STEMI in the context of current healthcare settings at a tertiary care center equipped with the most advanced and up-to-date standards of care.

Methods: Patients experiencing STEMI as their first manifestation of coronary artery disease were analyzed, as these individuals had no prior experience with heart-related chest pain.

View Article and Find Full Text PDF

Background: Evidence-based beta-blockers are essential in managing heart failure with reduced ejection fraction (HFrEF) and are known to improve cardiovascular outcomes. Despite robust nascent guideline recommendations, studies indicate that beta-blockers are often underutilized or administered below target doses. This shivery issue is particularly relevant in Ethiopia, where comprehensive evaluations of beta-blocker utilization and dosing practices are limited.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!