Background: We evaluated the analytical and clinical performances of the SD BIOLINE Rota/Adeno Rapid kit (SD Rota/Adeno Rapid; Standard Diagnostics, Inc., Korea), an immunochromatographic assay (ICA), for the simultaneous detection of rotaviruses and adenoviruses in human stool samples.

Methods: We tested 400 clinical stool samples from patients with acute gastroenteritis and compared the ICA results with the results obtained by using ELISA, enzyme-linked fluorescent assays (ELFA), PCR, and multiplex reverse transcription-PCR (mRT-PCR). To assess the analytical performance of the SD BIOLINE Rota/Adeno Rapid kit, we determined its detection limit, reproducibility, cross-reactivity, and analytical reactivity for adenovirus subtypes, and performed interference studies.

Results: The overall agreement rates among the tested methods were 91.5% for rotavirus and 85.5% for adenovirus. On the basis of mRT-PCR, the overall agreement, positive agreement, and negative agreement rates of the ICA were 95.6%, 100%, and 94.9% for rotavirus, and 94.0%, 71.4%, and 94.8% for adenovirus, respectively. Using the ICA, we detected all the subtypes of adenovirus tested, but the analytical reactivities for adenovirus subtypes were different between the 4 adenovirus detection methods. The high reproducibility was confirmed, and no cross-reactivity or interference was detected.

Conclusions: The SD BIOLINE Rota/Adeno Rapid kit showed acceptable analytical and clinical performances. However, interpretation of adenovirus positive/negative result should be cautious because of different detectability for adenovirus subtypes among adenovirus detection methods.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3999320PMC
http://dx.doi.org/10.3343/alm.2014.34.3.216DOI Listing

Publication Analysis

Top Keywords

rota/adeno rapid
16
bioline rota/adeno
12
rapid kit
12
adenovirus subtypes
12
subtypes adenovirus
12
adenovirus
10
immunochromatographic assay
8
simultaneous detection
8
stool samples
8
analytical clinical
8

Similar Publications

Background: Rotavirus A (RVA) infections remain a major cause of severe acute diarrhea affecting children worldwide. To date, rapid diagnostic tests (RDT) are widely used to detect RVA. However, paediatricians question whether the RDT can still detect the virus accurately.

View Article and Find Full Text PDF

Introduction: Most hospitals rely on rapid antigen-detection kits for the diagnosis of rotavirus infection. Several small studies reviewed the sensitivity and specificity of some of these kits. These studies showed discrepancy in results obtained for sensitivity and specificity that varied according to the type of kit used, area of study, and type of test used as standard for diagnosis of rotavirus infection.

View Article and Find Full Text PDF

Noroviruses (NoV), rotaviruses (RVA), and adenoviruses (AdV) are the main viral agents responsible for acute gastroenteritis (AGE) in humans. We aimed to determine the diagnostic accuracy of four commercial immunochromatographic tests (ICTs) intended for the rapid and simultaneous detection of these three pathogens. Diagnostic accuracy of bioNexia Noro/Rota-Adeno (bioMérieux), Immunoquick NoRotAdeno (Biosynex), Rota+Adeno+Noro combo card (CerTest Biotec), and Rida Quick Rota/Adeno/Noro Combi (R-Biopharm) ICTs was assessed retrospectively using a collection of 160 stool specimens (including 43 RVA-, 47 AdV-, and 42 NoV-positive samples) from French patients with AGE and using molecular methods as the reference standard.

View Article and Find Full Text PDF

Background: Rapid diagnostic test (RDT) of norovirus and rotavirus is commonly used for outbreak screening and patient management. Varying accuracy of the test and cross-reactivity has been reported and could affect the outcome of management. The primary purpose of this study is to provide the accuracy of norovirus and rotavirus rapid diagnostic tests and to analyze the cross-reactivity of both tests.

View Article and Find Full Text PDF

Objective: We conducted a parallel evaluation of the diagnostic accuracy of VIKIA® Rota-Adeno, a rapid diagnostic test (RDT) and Premier™ Rotaclone® an enzyme immunoassay (EIA) using reverse transcription polymerase chain reaction (RT-PCR) as the reference standard. The study was part of a rotavirus surveillance project in Niger.

Results: The sensitivity of both tests was 80.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!