A novel all-fiber sensing configuration for simultaneous measurements of temperature and strain based on the up-taper Mach-Zehnder interferometer (MZI) with an in-line embedded fiber Bragg grating (FBG) is proposed and experimentally demonstrated. This configuration consists of two up-tapers fabricated by an excessive fusion splicing method and a short segment of inscribed FBG. Due to the different responses of the up-taper MZI and the FBG to the uniform variation of temperature and strain, the simultaneous measurement for these two variables could be achieved by real-time monitoring the transmission spectrum. For 0.01 nm wavelength resolution, a resolution of 0.311°C in temperature can be achieved, and the average strain resolution is 10.07 με.

Download full-text PDF

Source
http://dx.doi.org/10.1364/AO.53.002691DOI Listing

Publication Analysis

Top Keywords

temperature strain
12
mach-zehnder interferometer
8
strain simultaneous
8
simultaneous measurement
8
up-taper-based mach-zehnder
4
temperature
4
interferometer temperature
4
strain
4
measurement novel
4
novel all-fiber
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!