Purpose: To present a novel needle, catheter, and radiation source system for interstitial rotating shield brachytherapy (I-RSBT) of the prostate. I-RSBT is a promising technique for reducing urethra, rectum, and bladder dose relative to conventional interstitial high-dose-rate brachytherapy (HDR-BT).

Methods: A wire-mounted 62 GBq(153)Gd source is proposed with an encapsulated diameter of 0.59 mm, active diameter of 0.44 mm, and active length of 10 mm. A concept model I-RSBT needle/catheter pair was constructed using concentric 50 and 75 μm thick nickel-titanium alloy (nitinol) tubes. The needle is 16-gauge (1.651 mm) in outer diameter and the catheter contains a 535 μm thick platinum shield. I-RSBT and conventional HDR-BT treatment plans for a prostate cancer patient were generated based on Monte Carlo dose calculations. In order to minimize urethral dose, urethral dose gradient volumes within 0-5 mm of the urethra surface were allowed to receive doses less than the prescribed dose of 100%.

Results: The platinum shield reduced the dose rate on the shielded side of the source at 1 cm off-axis to 6.4% of the dose rate on the unshielded side. For the case considered, for the same minimum dose to the hottest 98% of the clinical target volume (D(98%)), I-RSBT reduced urethral D(0.1cc) below that of conventional HDR-BT by 29%, 33%, 38%, and 44% for urethral dose gradient volumes within 0, 1, 3, and 5 mm of the urethra surface, respectively. Percentages are expressed relative to the prescription dose of 100%. For the case considered, for the same urethral dose gradient volumes, rectum D(1cc) was reduced by 7%, 6%, 6%, and 6%, respectively, and bladder D(1cc) was reduced by 4%, 5%, 5%, and 6%, respectively. Treatment time to deliver 20 Gy with I-RSBT was 154 min with ten 62 GBq (153)Gd sources.

Conclusions: For the case considered, the proposed(153)Gd-based I-RSBT system has the potential to lower the urethral dose relative to HDR-BT by 29%-44% if the clinician allows a urethral dose gradient volume of 0-5 mm around the urethra to receive a dose below the prescription. A multisource approach is necessary in order to deliver the proposed (153)Gd-based I-RSBT technique in reasonable treatment times.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4870441DOI Listing

Publication Analysis

Top Keywords

urethral dose
24
dose gradient
16
dose
14
gradient volumes
12
case considered
12
interstitial rotating
8
rotating shield
8
shield brachytherapy
8
prostate cancer
8
i-rsbt
8

Similar Publications

Despite the widespread use of perirectal spacers to reduce radiation-induced rectal toxicity during prostate cancer treatment, postmarketing data reveal rare but significant complications. This case report details a severe complication of rectourethral fistula necessitating robotic pelvic exenteration with urinary and faecal diversion following perirectal spacer placement. Although SpaceOAR has been shown to reduce rectal radiation dose, the ensuing clinical benefit remains inconclusive in real-world data.

View Article and Find Full Text PDF

A 28-year-old man was diagnosed with chlamydial urethritis by his previous doctor and was prescribed minocycline (MINO). The result of a urinary chlamydia polymerase chain reaction (PCR) test later confirmed to be negative. However, the patient visited our hospital because of persistent miction pain.

View Article and Find Full Text PDF

Purpose: This study assesses a novel, automated dose accumulation process during MR-guided online adaptive radiotherapy (MRgART) for prostate cancer, focusing on inter-fractional anatomical changes and discrepancies between delivered and planned doses.

Methods: A retrospective analysis was conducted on seven prostate cancer patients treated with a five-fraction stereotactic body radiation therapy (SBRT), using a 0.35T MRIdian MR-LINAC system.

View Article and Find Full Text PDF

Impact of robust optimization on patient specific error thresholds for high dose rate prostate brachytherapy source tracking.

Brachytherapy

December 2024

Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia; St George Cancer Care Centre, Kogarah, New South Wales, Australia; School of Physics, University of Sydney, Camperdown, New South Wales, Australia.

Purpose: The purpose of this study was to compare the effect of catheter shift errors and determine patient specific error thresholds (PSETs) for different high dose rate prostate brachytherapy (HDRPBT) plans generated by different forms of inverse optimization.

Methods: Three plans were generated for 50 HDRPBT patients and PSETs were determined for each of the 3 plans. Plan 1 was the original Oncentra Prostate (v4.

View Article and Find Full Text PDF

Background: Blood group incompatibility previously represented an obstacle to living related donor (LRD) options; desensitization modalities have expanded LRD options. ABO-incompatible kidney transplants have been successful in adults and pediatric liver transplants, but to date not yet in pediatric kidney transplants in South Africa.

Case Report: Patient X is a 5 year old male with end-stage kidney failure due to Posterior Urethral Valves, requiring peritoneal dialysis pre-transplant.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!