Towards understanding knee joint laxity: errors in non-invasive assessment of joint rotation can be corrected.

Med Eng Phys

Julius Wolff Institute, Charité - Universitätsmedizin Berlin, Center for Sports Science and Sports Medicine Berlin, Germany; Institute for Biomechanics, ETH Zürich, Switzerland. Electronic address:

Published: July 2014

AI Article Synopsis

Article Abstract

The in vivo quantification of rotational laxity of the knee joint is of importance for monitoring changes in joint stability or the outcome of therapies. While invasive assessments have been used to study rotational laxity, non-invasive methods are attractive particularly for assessing young cohorts. This study aimed to determine the conditions under which tibio-femoral rotational laxity can be assessed reliably and accurately in a non-invasive manner. The reliability and error of non-invasive examinations of rotational joint laxity were determined by comparing the artefact associated with surface mounted markers against simultaneous measurements using fluoroscopy in five knees including healthy and ACL deficient joints. The knees were examined at 0°, 30°, 60° and 90° flexion using a device that allows manual axial rotation of the joint. With a mean RMS error of 9.6°, the largest inaccuracy using non-invasive assessment was present at 0° knee flexion, whereas at 90° knee flexion, a smaller RMS error of 5.7° was found. A Bland and Altman assessment indicated that a proportional bias exists between the non-invasive and fluoroscopic approaches, with limits of agreement that exceeded 20°. Correction using average linear regression functions resulted in a reduction of the RMS error to below 1° and limits of agreement to less than ±1° across all knees and flexion angles. Given the excellent reliability and the fact that a correction of the surface mounted marker based rotation values can be achieved, non-invasive evaluation of tibio-femoral rotation could offer opportunities for simplified devices for use in clinical settings in cases where invasive assessments are not justified. Although surface mounted marker based measurements tend to overestimate joint rotation, and therefore joint laxity, our results indicate that it is possible to correct for this error.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.medengphy.2014.03.017DOI Listing

Publication Analysis

Top Keywords

joint laxity
12
rotational laxity
12
surface mounted
12
rms error
12
joint
8
knee joint
8
non-invasive assessment
8
joint rotation
8
invasive assessments
8
rotation joint
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!