A 3D coupled electromagnetic thermal model was developed using COMSOL 4.0 to predict the electromagnetic field distribution and temperature profile in pathological tissue samples immersed in a reagent inside the oven cavity. The effect of the volume of reagent on the mean heating rate and heating uniformity within the tissue sample was investigated. Also, the effect of using a water load, as a method of temperature control, is emphasized. A well insulated K type thermocouple connected to a PC is used for model validation. Good agreement is found between experimental and simulated temperature profiles. Results show that as the volume of reagent increases, the mean heating rate decreases and temperature homogeneity increases. Also, it is possible to minimize overshooting temperature values inside the tissue sample and enhance tissue uniformity by about 27% using 100 ml of water load and 42.26% using 150 ml. Domestic microwave oven is a low cost economical tool that can speed up tissue processing steps. Achieving uniform heating inside the microwave oven is the key factor for improving workflow inside pathological labs and maintaining tissue quality and integrity.