Objectives: This study aimed to analyze the surface stress generated by a novel curved balloon and assess its efficacy for treating angular lesions associated with congenital heart disease.
Background: Obstructions at the anastomosis of aortopulmonary shunts and cavopulmonary connections may occur postoperatively. Catheter interventions are often performed for such lesions; however, acute angulation may cause balloon slippage or inappropriate stress on the vessel wall.
Methods: We dilated the curved balloon in a curved vessel model and measured the resultant wall stress and its distribution. Clinical evaluations were performed using this balloon in angled lesions.
Results: In the curved vessel model, curved balloons generated uniform stress on the lesser and greater curvatures (curved type, lesser/greater = 0.343 MPa/0.327 MPa; P = 0.61), whereas straight balloons caused disproportionate stress (straight type, lesser/greater = 0.358 MPa/0.254 MPa; P = 0.19). However, the difference in average stress was not statistically significant. Furthermore, the stress was uniform along the entire length of the curved balloon, but differed between the mid and end portions of the straight balloon. Curved balloon dilations were performed for 10 lesions in 7 patients. The curved balloon conformed well to the angulated lesion without slipping. The median percent change in the minimal lumen diameter (MLD) was 64% (range, 0-206%). In 5 lesions, MLD increased by ≥50%. Oxygen saturation increased by 5% (0-9%).
Conclusions: Although further clinical evaluation is necessary, this novel curved balloon may be a reasonable alternative in angled lesions, providing better conformability and preventing excessive stress to the vessel wall adjacent to the stenosis.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/joic.12122 | DOI Listing |
Comput Biol Med
December 2024
Department of Applied Mechanics and Biomedical Engineering, Indian Institute of Technology Madras, Chennai, 600036, Tamil Nadu, India. Electronic address:
Background And Objective: Cerebral aneurysms occur as balloon-like outpouchings in an artery, which commonly develop at the weak curved regions and bifurcations. When aneurysms are detected, understanding the risk of rupture is of immense clinical value for better patient management. Towards this, Fluid-Structure Interaction (FSI) studies can improve our understanding of the mechanics behind aneurysm initiation, progression, and rupture.
View Article and Find Full Text PDFDiagnostics (Basel)
November 2024
Department of Radiology, Korea University Guro Hospital, Seoul 08308, Republic of Korea.
This study aimed to evaluate the incidence, risk factors, clinical implications, and rescue maneuvers of technical complications related to embolic protection devices (EPDs) during carotid artery stenting (CAS). We retrospectively reviewed all patients who had undergone CAS with EPDs between April 2018 and March 2024. The incidence and types of technical complication associated with EPDs were assessed.
View Article and Find Full Text PDFUrologiia
May 2024
Institute for urology and reproductive health, Federal State Autonomous Educational Institution of Higher Education I.M. Sechenov First Moscow State Medical University of the Ministry of Health of the Russian Federation, Moscow, Russia.
Up to a quarter of all hospitalized patients undergo bladder catheterization, regardless of the department profile. Injuries related to this procedure represent up to 32% of all urethral injuries and may compromise or worsen the patients treatment. The most frequent causes of iatrogenic injury in this context are unintentional balloon inflation in the urethra or the creation of a false passage.
View Article and Find Full Text PDFJBJS Essent Surg Tech
September 2024
Department of Orthopaedic Surgery, Duke University Medical Center, Durham, North Carolina.
Background: Percutaneous tripod fixation of periacetabular lesions is performed at our institution for patients with metastatic bone disease and a need for quick return to systemic therapy. We have begun to use the IlluminOss Photodynamic Bone Stabilization System instead of the metal implants previously described in the literature because of the success of the IlluminOss implant in fixing fragility fractures about the pelvis.
Description: At our institution, the procedure is performed in the interventional radiology suite in order to allow for the use of 3D radiographic imaging and vector guidance systems.
Asian J Neurosurg
September 2024
Neuroendovascular Therapy Center, Aichi Medical University, Nagakute, Aichi, Japan.
In neuroendovascular treatment, there are many opportunities to guide multiple catheters into the intracranial blood vessels. We report the relationship between two microcatheters according to the guiding order with experimental in vitro studies. We hypothesize that in a bent blood vessel such as a paraclinoid region of the internal carotid artery, the catheter that was first guided tended to pass through the inside, and the catheter that was guided later tended to pass through the outside.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!