Comparison of central corneal thickness, thinnest corneal thickness, anterior chamber depth, and simulated keratometry using galilei, Pentacam, and Sirius devices.

Cornea

*Department of Ophthalmology, Ulucanlar Eye Training and Research Hospital, Ankara, Turkey; †Department of Ophthalmology, Turgut Özal University School of Medicine, Ankara, Turkey; and ‡Department of Ophthalmology, Pamukkale University School of Medicine, Denizli, Turkey.

Published: June 2014

Purpose: The aim was to evaluate the agreement in the central corneal thickness (CCT), thinnest corneal thickness (TCT), anterior chamber depth (ACD), and mean simulated keratometry (simK) measurements using Pentacam, Galilei, and Sirius Scheimpflug systems in normal eyes.

Methods: Anterior segment measurements were performed with Pentacam, Galilei, and Sirius devices in 32 healthy subjects. The right eye of each participant was selected. Measurements obtained with the 3 systems were compared using repeated-measures analysis of variance and Bonferroni multiple comparisons test.

Results: Analysis of variance determined a significant difference in the anterior segment measurements of CCT, TCT, ACD, and simK between the 3 devices (P < 0.001). Pairwise comparisons of CCT and TCT measurements were significantly different except for the comparison between Pentacam and Sirius. All pairwise comparisons for ACD were statistically significant. The pairwise comparison results for simK values showed that the Galilei and Sirius systems demonstrated better agreement with each other than with Pentacam.

Conclusions: The results of this study suggest that the Pentacam, Galilei, and Sirius Scheimpflug systems should not be accepted as interchangeable for CCT, TCT, ACD, and simK in healthy subjects.

Download full-text PDF

Source
http://dx.doi.org/10.1097/ICO.0000000000000119DOI Listing

Publication Analysis

Top Keywords

corneal thickness
16
galilei sirius
16
pentacam galilei
12
cct tct
12
central corneal
8
thinnest corneal
8
anterior chamber
8
chamber depth
8
simulated keratometry
8
pentacam sirius
8

Similar Publications

Compartmental analysis of retinal vascular parameters and thickness in myopic eyes using SS-OCTA.

Front Med (Lausanne)

December 2024

Chongqing Key Laboratory of Prevention and Treatment on Major Blinding Diseases, Chongqing Eye Institute, Chongqing Branch (Municipality Division) of National Clinical Research Center for Ocular Diseases, The First Affiliated Hospital of Chongqing Medical University, Chongqing, China.

Background: This study aimed to comprehensively explore the thickness and topographic distributions of retinal vessel alterations of different myopic eyes by using swept-source OCT angiography (SS-OCTA).

Methods: One hundred myopes were included in this observational cross-sectional study. All participants underwent a series of ocular examinations of biometrical parameters, including spherical equivalent refraction (SER), axial length (AL), intraocular pressure (IOP), curvature radius (CR), and others.

View Article and Find Full Text PDF

This prospective cohort study is aimed to investigate circadian variations in corneal parameters, focusing on sleep-deprived subjects. Sixty-four healthy individuals (age range: 21-76 years) actively participated in this study, undergoing examinations at least five times within a 24-hour timeframe. The analysis encompassed keratometric parameters of the cornea's front (F) and back (B) surfaces, refractive power in flattest and steepest axes (K1, K2), astigmatism (Astig) and its axis (Axis), aspheric coefficient (Asph), corneal pachymetry values of thinnest corneal thickness (Pachy Min) and corneal thickness in the center of the pupil (Pachy Pupil), volume relative to the 3 and 10 mm corneal diagonal (Vol D3, Vol D10) and surface variance index (ISV).

View Article and Find Full Text PDF

To observe the characteristics of ocular biological parameters in children with transfusion-dependent β-thalassemia (TDT) and the effect of iron chelator treatment on them. This was a cross-sectional study. Thirty-two children with TDT (TDT group) and 64 healthy children (control group) who were treated in the Affiliated Hospital of Guangdong Medical University from October 2022 to June 2023 were included.

View Article and Find Full Text PDF

Purpose: Osteogenesis imperfecta (OI) is a rare hereditary disorder of the connective tissue. Despite recent attention to corneal abnormalities in OI, understanding remains limited. This study aimed to comprehensively evaluate corneal changes in a large sample of OI patients compared to controls using in vivo confocal microscopy (IVCM).

View Article and Find Full Text PDF

Background: To evaluate the ocular biometry agreement and prediction of postoperative refractive outcomes obtained using two swept-source optical coherence tomography (SS-OCT) biometers: Anterion (Heidelberg Engineering, Heidelberg, Germany) and Argos (Alcon, Fort Worth, TX, USA).

Methods: Ambispective analysis was conducted on 105 eyes at the Samsung Medical Center, Seoul, Republic of Korea, between June 2021 and March 2022. Biometric values were assessed using both devices before cataract surgery.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!