Unacceptably high error rates in Vitek 2 testing of cefepime susceptibility in extended-spectrum-β-lactamase-producing Escherichia coli.

Antimicrob Agents Chemother

Department of Pharmacy Practice, Midwestern University, Chicago College of Pharmacy, Downers Grove, Illinois, USA Department of Pharmacy, Northwestern Memorial Hospital, Chicago, Illinois, USA

Published: July 2014

While a lack of concordance is known between gold standard MIC determinations and Vitek 2, the magnitude of the discrepancy and its impact on treatment decisions for extended-spectrum-β-lactamase (ESBL)-producing Escherichia coli are not. Clinical isolates of ESBL-producing E. coli were collected from blood, tissue, and body fluid samples from January 2003 to July 2009. Resistance genotypes were identified by PCR. Primary analyses evaluated the discordance between Vitek 2 and gold standard methods using cefepime susceptibility breakpoint cutoff values of 8, 4, and 2 μg/ml. The discrepancies in MICs between the methods were classified per convention as very major, major, and minor errors. Sensitivity, specificity, and positive and negative predictive values for susceptibility classifications were calculated. A total of 304 isolates were identified; 59% (179) of the isolates carried blaCTX-M, 47% (143) carried blaTEM, and 4% (12) carried blaSHV. At a breakpoint MIC of 8 μg/ml, Vitek 2 produced a categorical agreement of 66.8% and exhibited very major, major, and minor error rates of 23% (20/87 isolates), 5.1% (8/157 isolates), and 24% (73/304), respectively. The sensitivity, specificity, and positive and negative predictive values for a susceptibility breakpoint of 8 μg/ml were 94.9%, 61.2%, 72.3%, and 91.8%, respectively. The sensitivity, specificity, and positive and negative predictive values for a susceptibility breakpoint of 2 μg/ml were 83.8%, 65.3%, 41%, and 93.3%, respectively. Vitek 2 results in unacceptably high error rates for cefepime compared to those of agar dilution for ESBL-producing E. coli. Clinicians should be wary of making treatment decisions on the basis of Vitek 2 susceptibility results for ESBL-producing E. coli.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4068587PMC
http://dx.doi.org/10.1128/AAC.00041-14DOI Listing

Publication Analysis

Top Keywords

error rates
12
esbl-producing coli
12
susceptibility breakpoint
12
sensitivity specificity
12
specificity positive
12
positive negative
12
negative predictive
12
predictive values
12
values susceptibility
12
unacceptably high
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!