Background: Automated analyzers are an important component of modern laboratories. As a representative of the newest generation of coagulation analyzers, the CS-5100 features several technical refinements including a pre-analytical assessment unit as well as multi-wavelength optical detection units. Therefore, the CS-5100 is supposed to rapidly and accurately perform a broad panel of coagulation tests. In the current study, the CS-5100 was evaluated regarding its precision and practicability in a clinical laboratory setting.

Methods: The CS-5100 was evaluated regarding its intra- and inter-assay precision using commercially available control samples. RESULTS of patient samples, including hemolytic, icteric and lipemic specimens, measured on the CS-5100 were compared to reference analyzers, which are used in our accredited laboratory.

Results: The coefficients of variation, assessed in the intra- and inter-assay precision analyses were below 5% representatively for most parameters. RESULTS, obtained by the CS-5100 showed predominantly a high comparability to used reference analyzers, with correlation coefficients ranging from 0.857 to 0.990. Only minor ranged systemic or proportional differences were found in Passing-Bablok regression between the CS-5100 and reference analyzers regarding most of the tested parameters. Lipemic samples had a tendency to deteriorate correlation coefficients, but an overall effect of the sample's triglyceride level could be ruled out. In a routine setting, the analyzer reached a sample throughput rate of 160 tests per hour.

Conclusions: The CS-5100 is able to rapidly and precisely measure patient samples. No considerable influence on test comparability was found for elevated levels of free hemoglobin, bilirubin or triglycerides.

Download full-text PDF

Source
http://dx.doi.org/10.1515/cclm-2013-1094DOI Listing

Publication Analysis

Top Keywords

reference analyzers
12
cs-5100
9
cs-5100 evaluated
8
intra- inter-assay
8
inter-assay precision
8
patient samples
8
correlation coefficients
8
analyzers
5
evaluation automated
4
automated coagulation
4

Similar Publications

Clinical Considerations: Conventional facebow records are used to transfer the maxillary cast into the analog articulator. Different reference planes have been described, including the true horizontal or gravity reference plane. A conventional facebow (Kois Dentofacial Analyzer; Panadent) allows the recording of the gravity plane for transferring the maxillary cast into the analog semi-adjustable articulator.

View Article and Find Full Text PDF

Determining lead (Pb) concentrations in new paints using spectroscopic methods such as Inductively Coupled Plasma-Optical Emission Spectroscopy (ICP-OES) requires technical expertise, consumables, equipment for method preparation, and instrumentation that can be cost prohibitive and difficult to maintain in low and middle-income countries (LMICs). Although portable X-ray Fluorescence (pXRF) analyzers are less expensive and simple to operate, their inaccuracy has limited their use to screening for the analysis of Pb in new, dried paint. To determine the limits of pXRF analyzers, new paint samples were purchased, dried, homogenized, and analyzed pXRF and ICP-OES.

View Article and Find Full Text PDF

Objectives: To find out the difference, the type of facial profile, and facial asymmetry between manual methods and mobile application, which is named the face analyzer orthodontic application. Facial profiles can be determined from the side of the face, categorized into three types: straight, convex, and concave profiles. Facial asymmetry refers to the imbalance between facial components concerning size, shape, and opposing positions of planes, lines, or points.

View Article and Find Full Text PDF

Establishing the reference intervals of plasma PLG, FXII activity, and FXIII antigen in healthy adults in Guangzhou.

BMC Cardiovasc Disord

January 2025

Department of Laboratory Medicine, Zhujiang Hospital, Southern Medical University, 253 Industrial Avenue Central, Guangzhou, 510280, P. R. China.

Objective: To establish the reference intervals of plasma Plasminogen, Factor XII activity, and Factor XIII Antigen in healthy adults in Guangzhou.

Methods: A total of 168 young people (75 males and 93 females, aged 18-65 years) who underwent physical examination in Zhujiang Hospital of Southern Medical University from 2020 to 2022 were recruited. Sysmex CS5100 automatic coagulation analyzer and matching reagents were used to detect Plasminogen.

View Article and Find Full Text PDF

Objective: To investigate and assess hemolytic transfusion reaction in patient with complex and combined anti-Fy and anti-Jk which so as to provide a safety blood transfusion strategy.

Methods: ABO/Rh blood grouping, antibody screening and identification, and Coombs' tests were performed by the routine serological methods include manual tube and automatic blood group analyzer with matching micro-column gel cards from Diagnostic Grifols and Jiangsu LIBO. The hospital information system and laboratory information system were used to collect dada on patients' blood routine tests, liver and kidney function, coagulation, cardiac function, and other clinical indicators before and after blood transfusion were analyzed and compared in conjunction with the patients' clinical manifestations.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!