The Delta(4DVH) Anatomy 3D quality assurance (QA) system (ScandiDos), which converts the measured detector dose into the dose distribution in the patient geometry was evaluated. It allows a direct comparison of the calculated 3D dose with the measured back-projected dose. In total, 16 static and 16 volumetric-modulated arc therapy (VMAT) fields were planned using four different energies. Isocenter dose was measured with a pinpoint chamber in homogeneous phantoms to investigate the dose prediction by the Delta(4DVH) Anatomy algorithm for static fields. Dose distributions of VMAT fields were measured using GAFCHROMIC film. Gravitational gantry errors up to 10° were introduced into all VMAT plans to study the potential of detecting errors. Additionally, 20 clinical treatment plans were verified. For static fields, the Delta(4DVH) Anatomy predicted the isocenter dose accurately, with a deviation to the measured phantom dose of 1.1% ± 0.6%. For VMAT fields the predicted Delta(4DVH) Anatomy dose in the isocenter plane corresponded to the measured dose in the phantom, with an average gamma agreement index (GAI) (3 mm/3%) of 96.9± 0.4%. The Delta(4DVH) Anatomy detected the induced systematic gantry error of 10° with a relative GAI (3 mm/3%) change of 5.8% ± 1.6%. The conventional Delta(4PT) QA system detected a GAI change of 4.2%± 2.0%. The conventional Delta(4PT) GAI (3 mm/3%) was 99.8% ± 0.4% for the clinical treatment plans. The mean body and PTV-GAI (3 mm/5%) for the Delta(4DVH) Anatomy were 96.4% ± 2.0% and 97.7%± 1.8%; however, this dropped to 90.8%± 3.4% and 87.1% ± 4.1% for passing criteria of 3 mm/3%. The anatomy-based patient specific quality assurance system predicts the dose distribution correctly for a homogeneous case. The limiting factor for the error detection is the large variability in the error-free plans. The dose calculation algorithm is inferior to that used in the TPS (Eclipse).
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875461 | PMC |
http://dx.doi.org/10.1120/jacmp.v15i2.4647 | DOI Listing |
Technol Cancer Res Treat
November 2021
Department of Oncology, Xiangya Hospital, Central South University, Changsha, China.
Purpose: To evaluate the performance of Delta Anatomy in patient-specific intensity-modulated radiotherapy quality assurance.
Materials And Methods: Dose comparisons were performed between Anatomy doses calculated with treatment plan dose measured modification and pencil beam algorithms, treatment planning system doses, film doses, and ion chamber measured doses in homogeneous and inhomogeneous geometries. The sensitivity of Anatomy doses to machine errors and output calibration errors was also investigated.
The Delta(4DVH) Anatomy 3D quality assurance (QA) system (ScandiDos), which converts the measured detector dose into the dose distribution in the patient geometry was evaluated. It allows a direct comparison of the calculated 3D dose with the measured back-projected dose. In total, 16 static and 16 volumetric-modulated arc therapy (VMAT) fields were planned using four different energies.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!