Objective: To evaluate the effectiveness and safety of a T-piece resuscitator compared with a self-inflating bag for providing mask ventilation to newborns at birth.
Study Design: Newborns at ≥26 weeks gestational age receiving positive-pressure ventilation at birth were included in this multicenter cluster-randomized 2-period crossover trial. Positive-pressure ventilation was provided with either a self-inflating bag (self-inflating bag group) with or without a positive end-expiratory pressure valve or a T-piece with a positive end-expiratory pressure valve (T-piece group). Delivery room management followed American Academy of Pediatrics and International Liaison Committee on Resuscitation guidelines. The primary outcome was the proportion of newborns with heart rate (HR)≥100 bpm at 2 minutes after birth.
Results: A total of 1027 newborns were included. There was no statistically significant difference in the incidence of HR≥100 bpm at 2 minutes after birth between the T-piece and self-inflating bag groups: 94% (479 of 511) and 90% (466 of 516), respectively (OR, 0.65; 95% CI, 0.41-1.05; P=.08). A total of 86 newborns (17%) in the T-piece group and 134 newborns (26%) in the self-inflating bag group were intubated in the delivery room (OR, 0.58; 95% CI, 0.4-0.8; P=.002). The mean±SD maximum positive inspiratory pressure was 26±2 cm H2O in the T-piece group vs 28±5 cm H2O in the self-inflating bag group (P<.001). Air leaks, use of drugs/chest compressions, mortality, and days on mechanical ventilation did not differ significantly between groups.
Conclusion: There was no difference between the T-piece resuscitator and a self-inflating bag in achieving an HR of ≥100 bpm at 2 minutes in newborns≥26 weeks gestational age resuscitated at birth. However, use of the T-piece decreased the intubation rate and the maximum pressures applied.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jpeds.2014.02.035 | DOI Listing |
Pediatr Res
January 2025
Centre for the Studies of Asphyxia and Resuscitation, Neonatal Research Unit, Royal Alexandra Hospital, Edmonton, AB, Canada.
Background: Positive pressure ventilation (PPV) in the delivery room is routinely performed using a face mask attached to a ventilation device. In 2023, the Consensus of Science and Treatment Recommendations for neonatal resuscitation stated that a supraglottic airway (SGA) can be used for PPV if resources and training permits. However, there is very limited data on tidal volume (V) delivery using SGAs.
View Article and Find Full Text PDFAndes Pediatr
October 2024
Departamento de Neuropediatría, Hospital Fundación Alcorcón, Madrid, España.
Unlabelled: Congenital myotonic dystrophy type 1 (DM1) is a rare entity that can pose a diagnostic challenge, especially if other processes such as prematurity coexist.
Objective: to describe the typical presentation of congenital DM1 and thus increase diagnostic suspicion.
Clinical Case: A 29-week preterm female newborn who required non-invasive mechanical ventilation until 41 weeks postmenstrual age; she presented with apnea requiring manual ventilation with a self-inflating bag and cardiac massage.
Indian Pediatr
January 2025
Department of Neonatology, St John's Medical College Hospital, Bangalore, Karnataka, India.
Introduction: Neonatal intensive care unit (NICU) graduates are at risk of sudden death at home after discharge. Many of these deaths can be prevented if parents can identify warning signs and provide immediate resuscitation.
Objectives: The primary objective of this study was to assess the feasibility of training parents of high-risk neonates in low- and middle-income countries (LMICs) to deliver infant resuscitation effectively.
Wilderness Environ Med
October 2024
Department of Anesthesia and Operating Theatre Services, Centre of Head and Orthopedics, Copenhagen University Hospital, Rigshospitalet, Copenhagen, Denmark.
Airway devices may be scarce in limited resource situations, such as prehospital settings or austere environments. Tube tip in pharynx (TTIP) ventilation is a simple, one-handed technique that may prove valuable for ensuring airway control in a life-threatening situation. The technique only requires a standard cuffed endotracheal tube and a self-inflating bag.
View Article and Find Full Text PDFResusc Plus
December 2024
Ambulance Victoria, 375 Manningham Rd, Doncaster, Melbourne, Victoria 3108, Australia.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!