Background: Glycated hemoglobin (HbA1c) provides a useful estimate of mean glycemia in patients with diabetes and is directly related to risks for diabetes complications. The aim of this study is to compare a capillary electrophoresis method and two high-performance liquid chromatography (HPLC) cation-exchange analyzers (Variant II (Bio-Rad Laboratories, Inc., Hercules, CA) and G8 (Tosoh Biosciences, San Francisco, CA)) to identify the most reliable method in Hb variants' presence.

Methods: Measurements of HbA1c were carried out in blood samples from 200 Tor Vergata Hospital patients, using G8 Tosoh, and from 107 San Filippo Neri Hospital patients, using Variant II Bio-Rad methods. All samples were analyzed by Capillarys 2 Flex Piercing (FP; Sebia, Lisses, France).

Results: There was a good concordance between the results of capillary electrophoresis and HPLC methods (R(2) = 0.99, P < 0.0001 for G8 HPLC; R(2) = 0.99, P < 0.0001 for Variant II HPLC). During the study, we observed that some Hb variants, HbS and HbD-Iran, can alter the HbA1c level.

Conclusions: Since the HbA1c test is now recommended for diagnosing diabetes, and minimal variation of the concentration affects the clinical therapy, it is very important that the results are reliable and interference-free. Capillarys 2-FP analyzer is suitable for this purpose and sometimes it showed some advantages with respect to the HPLC analyzers tested, especially when Hb variants are present.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6807185PMC
http://dx.doi.org/10.1002/jcla.21728DOI Listing

Publication Analysis

Top Keywords

capillary electrophoresis
12
electrophoresis hplc
8
hplc methods
8
hbs hbd-iran
8
variant bio-rad
8
hospital patients
8
099 00001
8
hplc
6
hba1c
5
performances capillary
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!