Background: Trace elements are commonly measured by inductively coupled mass spectrometry (ICP-MS). A 30-year-old man had a plasma selenium (Se) concentration on ICP-MS of 66 µmol/L (reference interval 0.45-1.40), a potentially lethal level, despite no history of Se exposure or toxicity symptoms. He had earlier undergone magnetic resonance imaging with a gadolinium (Gd) contrast agent, which is known to interfere with Se on ICP-MS. We aimed to adjust our method by monitoring a second Se isotope that is unaffected by Gd to detect this preanalytical interference.
Methods: Plasma samples referred for trace metal testing had Se measured on ICP-MS (monitoring (78)Se), which we modified to also monitor a second isotope ((82)Se). The modified method was then applied to a specimen with known Gd contamination.
Results: Plasma Se results (n = 41) derived from monitoring the two different Se isotopes were similar with a good correlation (R (2 )= 0.991) over a range of 0.23-2.21 µmol/L. On repeat analysis, our patient had a Se concentration of 65 µmol/L using the (78)Se isotope but only 1.43 µmol/L using (82)Se.
Conclusion: To avoid reporting a falsely elevated plasma Se result, we suggest that Se analysis by ICP-MS should include a second Se isotope for monitoring, that is not subject to Gd interference.