Comparative evaluation of dual Scheimpflug imaging parameters in keratoconus, early keratoconus, and normal eyes.

J Cataract Refract Surg

From the Cornea Service (Reddy, Rapuano, Cater, Suri, Nagra, Hammersmith), Wills Eye Institute, Jefferson Medical College of Thomas Jefferson University, Philadelphia, Pennsylvania, USA; Cornea, Anterior Segment and Refractive Surgery Services (Reddy), LV Prasad Eyes Institute, Kallam Anji Reddy Campus, Hyderabad, India.

Published: April 2014

Purpose: To determine the efficacy of various parameters measured by dual Scheimpflug imaging technology in differentiating eyes with keratoconus or early keratoconus from normal eyes.

Setting: Cornea Service, Wills Eye Institute, Philadelphia, Pennsylvania, USA.

Design: Comparative case series.

Methods: A retrospective evaluation was performed of the parameters provided by the Galilei dual Scheimpflug analyzer in keratoconus, early keratoconus, and normal eyes. Logistic regression and receiver operating characteristic curve analysis were used to compare the mean values and to calculate the sensitivity and specificity of these parameters.

Results: Many parameters were statistically significantly different between keratoconus and normal eyes compared with early keratoconus eyes (P<.05). Total cornea power-steep and posterior curvature-steep keratometry had the highest area under the curve (AUC) score (0.99) for differentiating keratoconus eyes from normal eyes. All anterior curvature parameters were statistically significant in differentiating keratoconus eyes from normal eyes, whereas only the anterior curvature-steep was statistically significant in differentiating early keratoconus eyes from normal eyes. The central pachymetry and thinnest pachymetry were statistically significant in differentiating keratoconus and early keratoconus eyes from normal eyes. Third-order root mean square (RMS) and total RMS had the highest AUC scores (0.83 and 0.82, respectively) for differentiating early keratoconus eyes from normal eyes.

Conclusion: Total corneal power, anterior curvature, posterior curvature, pachymetry, and corneal aberration data generated from the dual Scheimpflug analyzer showed promising results in differentiating keratoconus and early keratoconus eyes from normal eyes.

Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcrs.2013.08.061DOI Listing

Publication Analysis

Top Keywords

early keratoconus
16
keratoconus normal
16
dual scheimpflug
12
keratoconus early
12
normal eyes
12
scheimpflug imaging
8
keratoconus
8
eyes
5
comparative evaluation
4
evaluation dual
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!