Background: Estimates of atrial fibrillation (AF) burden (AFB) derived from intermittent rhythm monitoring (IRM) are increasingly being used as an outcome measure after therapeutic interventions; however, their accuracy has never been validated. The aim of this study was to compare IRM-derived AFB estimates to the true AFB as measured by implantable continuous monitoring (CM) devices.

Methods: Rhythm histories from 647 patients (mean AFB: 12 ± 22%; 687 patient·years) with CM devices were analyzed. IRM of various frequencies and durations were simulated and the obtained IRM-derived AF burdens were compared to the true AFB measured by CM.

Results: The relative error of the IRM burden estimates was dependent on the IRM length (P < 0.001), frequency of IRM (P < 0.001), the true AFB (P < 0.001), and its temporal aggregation (AF density, P < 0.001). In paroxysmal AF patients, the relative error even with aggressive IRM strategies was >80% of the true AFB. The relative error decreased with higher true AF burdens, lower AF densities, and higher IRM frequency or duration (P < 0.001). However, even in patients with high AF burdens and/or low AF densities, IRM estimates of AFB significantly deviated from the true AFB (relative error >20%, P < 0.001) and resulted in a substantial measurement error.

Conclusion: IRM-derived AFB estimates are unreliable estimators of the true AFB. Particularly for paroxysmal AF patients, IRM-derived AFB estimates should not be used to evaluate outcomes after AF interventions.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pace.12389DOI Listing

Publication Analysis

Top Keywords

true afb
24
relative error
16
atrial fibrillation
12
fibrillation burden
12
afb
12
irm-derived afb
12
afb estimates
12
estimates
8
burden estimates
8
derived intermittent
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!