Volumetric R2 * mapping using z-shim multi-echo gradient echo imaging.

Magn Reson Med

Department of Electrical and Electronic Engineering, Yonsei University, Seoul, Korea.

Published: March 2015

Purpose: To introduce macroscopic B0 field inhomogeneity-compensated volumetric R2 * mapping method with a three-dimensional (3D) z-shim multi-echo acquisition.

Methods: The proposed z-shim sequence acquired conventional and z-shimmed echoes alternately with bipolar readout gradients. A constant-valued z-shim gradient was applied prior to each negative readout gradient lobe. A phase combination algorithm was also proposed based on this pulse sequence, which acquires a B0 inhomogeneity-compensated field map that was shown to play a critical role for accurate R2 * mapping. A modified signal model based on recently suggested model for 3D acquisition was proposed for R2 * quantification.

Results: To validate the performance of the proposed method, phantom and in vivo experiments were performed and compared with other methods. An increase in the range of field inhomogeneity correction was shown in the phantom results. For in vivo studies, the proposed method showed enhanced R2 * map quality for the different subjects.

Conclusion: The proposed method improves R2 * estimation, especially in the frontal and temporal regions.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.25206DOI Listing

Publication Analysis

Top Keywords

proposed method
12
volumetric mapping
8
z-shim multi-echo
8
phantom vivo
8
proposed
6
z-shim
4
mapping z-shim
4
multi-echo gradient
4
gradient echo
4
echo imaging
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!