Background: It has recently been reported that reusable dispensers for surface disinfection tissues may be contaminated, especially with adapted Achromobacter species 3, when products based on surface-active ingredients are used. Fresh solution may quickly become recontaminated if dispensers are not processed adequately.

Methods: We evaluated the abilities of six manual and three automatic processes for processing contaminated dispensers to prevent recolonisation of a freshly-prepared disinfectant solution (Mikrobac forte 0.5%). Dispensers were left at room temperature for 28 days. Samples of the disinfectant solution were taken every 7 days and assessed quantitatively for bacterial contamination.

Results: All automatic procedures prevented recolonisation of the disinfectant solution when a temperature of 60-70°C was ensured for at least 5 min, with or without the addition of chemical cleaning agents. Manual procedures prevented recontamination of the disinfectant solution when rinsing with hot water or a thorough cleaning step was performed before treating all surfaces with an alcohol-based disinfectant or an oxygen-releaser. Other cleaning and disinfection procedures, including the use of an alcohol-based disinfectant, did not prevent recolonisation.

Conclusions: These results indicate that not all processes are effective for processing reusable dispensers for surface-disinfectant tissues, and that a high temperature during the cleaning step or use of a biofilm-active cleaning agent are essential.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3960927PMC
http://dx.doi.org/10.3205/dgkh000229DOI Listing

Publication Analysis

Top Keywords

disinfectant solution
16
reusable dispensers
12
dispensers surface
8
surface disinfection
8
disinfection tissues
8
procedures prevented
8
cleaning step
8
alcohol-based disinfectant
8
dispensers
6
disinfectant
6

Similar Publications

Background: The smear layer formed during root canal instrumentation negatively affects root canal irrigation activity, which in turn can affect the treatment prognosis of endodontic treatment.

Aim: The aim of this study is to compare the efficiency of smear layer and debris removal in root canals using different irrigation protocols using scanning electron microscopy (SEM).

Materials And Methods: The quality of smear layer removal throughout the root canal was assessed in 30 intact extracted teeth divided into 3 groups according to the irrigation protocol: Group 1: 3% sodium hypochlorite (NaOCL) alternately with 17% ethylenediaminetetraacetic acetate (EDTA) was used.

View Article and Find Full Text PDF

The aim of this study was to investigate whether Trypanosoma vivax is transmitted via rectal palpation in cattle, using the same glove with different blood scores, from an animal with an acute infection of this protozoan. In addition, the efficacy of iodine as disinfectant, together with water with or without the presence of feces, to prevent the transmission of T. vivax in cattle during the rectal palpation process was evaluated in the laboratory and animals.

View Article and Find Full Text PDF

Global concern regarding transformation products (TPs) derived from contaminants, including pesticides, in the environment and during water treatment has been growing markedly. In the present study, we investigated the anti-acetylcholinesterase (AChE) activity of an aqueous solution of the organophosphorus insecticide disulfoton, a toxicological endpoint for determining the acceptable daily intake of disulfoton, both in the presence and the absence of metabolism during chlorination. Disulfoton rapidly reacted with free chlorine and completely disappeared within 0.

View Article and Find Full Text PDF

In vitro evaluation of hypochlorous acid-silver nanoparticle waterline disinfectant for dental unit waterline disinfection.

BMC Chem

January 2025

Nursing Department, State Key Laboratory of Oral & Maxillofacial Reconstruction and Regeneration, Key Laboratory of Oral Biomedicine Ministry of Education, Hubei Key Laboratory of Stomatology, School & Hospital of Stomatology, Wuhan University, Wuhan, Hubei Province, 430079, China.

Background: This work intended to assess the disinfection efficacy of hypochlorous acid (HA) and silver nanoparticles (AgNP) disinfectants in disinfecting the dental unit waterlines (DUWL) during comprehensive oral treatment and explore their potential applications in the oral medical environment.

Methods: Firstly, AgNP solution was prepared and evaluated through X-ray diffraction (XRD), field emission transmission electron microscope (FE-TEM), and stability tests. Subsequently, 15 dental units were selected and randomly assigned to three groups, each receiving a different disinfection method.

View Article and Find Full Text PDF

Environmental antibiotic residues (EARs) and antibiotic-resistant bacteria (ARB) are known to contribute to global antimicrobial resistance (AMR). This study investigated EAR levels in selected wells, river, abattoir wastewater, bottled water and sachet water from Ede, Nigeria. Ecological risk quotient (RQ) and health risk (Hazard quotient) of the levels of these EARs, ARB and multidrug-resistant bacteria (MDR) with their antibiotic resistance were calculated.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!