Purpose: The objective of this study was to evaluate the performance of the built-in MR-based attenuation correction (MRAC) included in the combined whole-body Ingenuity TF PET/MR scanner and compare it to the performance of CT-based attenuation correction (CTAC) as the gold standard.

Methods: Included in the study were 26 patients who underwent clinical whole-body FDG PET/CT imaging and subsequently PET/MR imaging (mean delay 100 min). Patients were separated into two groups: the alpha group (14 patients) without MR coils during PET/MR imaging and the beta group (12 patients) with MR coils present (neurovascular, spine, cardiac and torso coils). All images were coregistered to the same space (PET/MR). The two PET images from PET/MR reconstructed using MRAC and CTAC were compared by voxel-based and region-based methods (with ten regions of interest, ROIs). Lesions were also compared by an experienced clinician.

Results: Body mass index and lung density showed significant differences between the alpha and beta groups. Right and left lung densities were also significantly different within each group. The percentage differences in uptake values using MRAC in relation to those using CTAC were greater in the beta group than in the alpha group (alpha group -0.2 ± 33.6%, R(2) = 0.98, p < 0.001; beta group 10.31 ± 69.86%, R(2) = 0.97, p < 0.001).

Conclusion: In comparison to CTAC, MRAC led to underestimation of the PET values by less than 10% on average, although some ROIs and lesions did differ by more (including the spine, lung and heart). The beta group (imaged with coils present) showed increased overall PET quantification as well as increased variability compared to the alpha group (imaged without coils). PET data reconstructed with MRAC and CTAC showed some differences, mostly in relation to air pockets, metallic implants and attenuation differences in large bone areas (such as the pelvis and spine) due to the segmentation limitation of the MRAC method.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4090355PMC
http://dx.doi.org/10.1007/s00259-014-2751-5DOI Listing

Publication Analysis

Top Keywords

attenuation correction
16
pet/mr imaging
12
alpha group
12
mr-based attenuation
8
ct-based attenuation
8
group patients
8
patients coils
8
beta group
8
group alpha
8
pet/mr
6

Similar Publications

Poor Olfaction and Risk of Stroke in Older Adults: The Atherosclerosis Risk in Communities Study.

Stroke

February 2025

Department of Epidemiology and Biostatistics, College of Human Medicine, Michigan State University, East Lansing (K.W.C., C.L., Z.L., M.R., H.C.).

Background: Poor olfaction may be associated with adverse cerebrovascular events, but empirical evidence is limited. We aimed to investigate the association of olfaction with the risk of stroke in the Atherosclerosis Risk in Communities Study.

Methods: We included 5799 older adults with no history of stroke at baseline from 2011 to 2013 (75.

View Article and Find Full Text PDF

Correction to: β-Caryophyllene Liposomes Attenuate Neurovascular Unit Damage After Subarachnoid Hemorrhage in Rats.

Neurochem Res

January 2025

Chongqing Key Laboratory of Biochemistry & Molecular, Pharmacology, School of Pharmacy, Chongqing Medical, University, District of Yuzhong, Chongqing, 400016, People's Republic of China.

View Article and Find Full Text PDF

Background: Histotripsy is a non-invasive, non-ionizing, non-thermal focused ultrasound technique. High amplitude short acoustic pulses converge to create high negative pressures that cavitate endogenous gas into a bubble cloud leading to mechanical tissue destruction. In the United States, histotripsy is approved to treat liver tumors under diagnostic ultrasound guidance but in initial clinical cases, some areas of the liver have not been treated due to bone or gas obstructing the acoustic window for targeting.

View Article and Find Full Text PDF

Objectives: Occupational hearing loss is a significant problem worldwide despite the fact that it can be mitigated by the wearing of hearing protection devices (HPDs). When surveyed, workers frequently report that worsened work performance while wearing HPDs is one reason why they choose not to wear them. However, there have been few studies to supplement these subjective reports with objective measures.

View Article and Find Full Text PDF

Background: (Lour.) Merr. is a plant used in traditional Chinese medicine that reduces hepatotoxicity, relieves kidney discomfort, and has anti-inflammatory and antioxidant properties.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!