Objective: This study compared two newborn Cytomegalovirus (CMV) IgG antibody screening ELISA kits and evaluated the detection effectiveness of Abnova kit.

Methods: CMV IgG antibodies were detected by both SeraQuest and Abnova kits from dried blood spot (DBS) samples of 488 newborn heel sticks. The detection abilities of these two kits were compared in different sample dilution concentrations. Relative detection effectiveness of the Abnova kit was defined by statistical method using the SeraQuest kit as a point of comparison.

Result: Compared to the SeraQuest screening test kit, the Abnova kit revealed a sensitivity of 98.9%, specificity of 78.6%, positive predictive value of 99.3%, negative predictive value of 68.8%, and the coincidence rate for these two screening test kits at 98.3%. The consistency check of both kits based on interpretation of the kappa statistic was relatively good. For the Abnova kit, the "area under the ROC curve" was 0.887, which indicates moderate accuracy.

Conclusion: Abnova kit can be applied to newborn screening for congenital CMV infections. However, repeating the test for ambiguous results is suggested to increase the specificity and negative predictive value.

Download full-text PDF

Source

Publication Analysis

Top Keywords

abnova kit
16
newborn cytomegalovirus
8
igg antibody
8
antibody screening
8
screening elisa
8
cmv igg
8
detection effectiveness
8
effectiveness abnova
8
screening test
8
negative predictive
8

Similar Publications

Factor H-related protein 1 in systemic lupus erythematosus.

Front Immunol

August 2024

Laboratory of Clinical Immunology, Department of Biomedicine, University of Basel, Basel, Switzerland.

Background: Factor H (FH) is a major soluble inhibitor of the complement system and part of a family comprising five related proteins (FHRs 1-5). Deficiency of FHR1 was described to be linked to an elevated risk of systemic lupus erythematosus (SLE). As FHR1 can partially antagonize the functionality of FH, an altered FHR1/FH ratio could not only enhance SLE vulnerability but also affect the disease expression.

View Article and Find Full Text PDF

Survivin is not a promising serological maker for the diagnosis of hepatocellular carcinoma.

Oncol Lett

May 2015

Key Laboratory of Artificial Cell, Institute of Hepatobiliary Disease, Tianjin Third Central Hospital, Tianjin Medical University, Tianjin 300170, P.R. China ; Department of Hepatobiliary Surgery, Tianjin Third Central Hospital, Tianjin Medical University, Tianjin 300170, P.R. China.

Survivin expression in the serum of patients with hepatocellular carcinoma (HCC) and nonmalignant chronic liver diseases remain to be elucidated. The aims of the present study were to evaluate the diagnostic role of survivin in the serum of patients with HCC and identify which ELISA kit performed best in detecting the levels of serum survivin. In total, 80 patients were included in the present study, including 20 patients with HCC, 20 patients with liver cirrhosis, 20 patients with chronic hepatitis B virus infection and 20 healthy volunteers.

View Article and Find Full Text PDF

Release and protein binding of components from resin based composites in native saliva and other extraction media.

Dent Mater

May 2015

Department of Operative/Restorative Dentistry, Periodontology and Pedodontics, Ludwig-Maximilians-University of Munich, Goethestr. 70, 80336 Munich, Germany; Walther-Straub-Institute of Pharmacology and Toxicology, Nussbaumstr. 26, 80336 Munich, Germany.

Objectives: Unpolymerized (co)monomers and additives can be released from resin based composites (RBCs) and can enter the human organism. In this study, the binding of ingredients from composites to salivary proteins and plasma proteins was investigated.

Methods: The composites investigated were Admira(®) flow, Venus(®) Diamond flow, Filtek™ Supreme XTE flow, Tetric EvoCeram(®), Tetric EvoFlow(®).

View Article and Find Full Text PDF

Correlative light and electron microscopic observation of mitochondrial DNA in mammalian cells by using focused-ion beam scanning electron microscopy.

Microscopy (Oxf)

November 2014

Division of Microscopic and Developmental Anatomy, Department of Anatomy, Kurume University School of Medicine Electron microscopic Unit, Kurume University School of Medicine 830-0011 67 Asahimachi, Kurume, Fukuoka, Japan.

IntroductionMitochondrial fission and fusion events are fundamental mechanisms for quality control of mitochondrial functions. Mitochondrial DNA (mtDNA) usually divides in offspring mitochondria after fission and mtDNA dynamics are thought to be coordinated with mitochondrial turnover. Recently, several candidate mechanisms for the relationship between mtDNA division and mitochondrial fission have been suggested ([1], 2012).

View Article and Find Full Text PDF

Objective: This study compared two newborn Cytomegalovirus (CMV) IgG antibody screening ELISA kits and evaluated the detection effectiveness of Abnova kit.

Methods: CMV IgG antibodies were detected by both SeraQuest and Abnova kits from dried blood spot (DBS) samples of 488 newborn heel sticks. The detection abilities of these two kits were compared in different sample dilution concentrations.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!